COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1193004 · Received October 13, 2008

Report

Report Number
6000001-2007-03523
Event Type
Malfunction
Date Received
October 13, 2008
Date of Event
January 1, 2007
Report Date
January 9, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA JAN 30, 2007. EVALUATION SUMMARY: THE CUSTOMER REPORTED ISSUE WAS CONFIRMED DURING SERVICE. INSPECTION OF THE DEVICE FOUND THAT THE MAIN BATTERIES WERE DEPLETED WITH 19 BATTERY DISCHARGES BELOW ALARM THRESHOLD AND WERE THEREFORE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.CONTINUED FROM H9:6000001-2/25/05-004-C6000001-12/13/05-019-CCONTINUED FROM D10:SERVICE OF THE DEVICE WAS CONDUCTED ON-SITE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED AN INFUSION PUMP WITH DEPLETED BATTERIES. THIS ISSUE REPORTED TO HAVE OCCURRED DURING BIO-MED TESTING. THE HOSPITAL REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.