FDA Adverse Event Injury Summary report: N

NXSTAGE SYSTEM ONE

MDR report key: 1192415 · Received October 7, 2008

Report

Report Number
3003464075-2008-00468
Event Type
Injury
Date Received
October 7, 2008
Date of Event
September 10, 2008
Report Date
September 10, 2008
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
PMA / PMN Number
K050525
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE OPERATOR NOT PERFORMING RINSEBACK DUE TO AIR IN THE CIRCUIT. FACILITY STAFF ATTRIBUTED TO ALARMS TO ISSUES WITH THE PATIENT'S ACCESS AND NEEDLE STICK. THERE IS NO EVIDENCE OF A DEVICE MALFUNCTION. A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. FACILITY STAFF HAS BEEN NOTIFIED. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFORMATION WILL BE PROVIDED.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. ARTERIAL PRESSURE AND VENOUS AIR ALARMS OCCURRED AT THE START OF A ROUTINE HEMODIALYSIS TREATMENT. RINSEBACK WAS NOT PERFORMED DUE TO THE AMOUNT OF AIR IN THE CIRCUIT RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR-170 8067702

Patients

Seq Age Sex Outcome Treatment
1 70 YR Other