TAXUS EXPRESS2 PACLITAXEL-ELUTING CORONARY STENT SYSTEM
Report
- Report Number
- 2134265-2008-02948
- Event Type
- Injury
- Date Received
- October 7, 2008
- Date of Event
- November 14, 2005
- Report Date
- September 23, 2008
- Manufacturer
- BOSTON SCIENTIFIC
- Product Code
- NIQ
- PMA / PMN Number
- P030025
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DC, US
- Reporter Occupation
- ATTORNEY
Narratives
THE COMPLAINANT INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVAL; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFO FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
SAME CASE AS: 2134265-2008-02614. IT WAS REPORTED, THAT POST A CORONARY DRUG ELUTING STENTING TREATMENT PROCEDURE, CHEST PAIN AND HOSPITALIZATION OCCURRED. THE 70-80% STENOSED LESION BEING TREATED WAS LOCATED IN THE TORTUOUS, LIGHTLY CALCIFIED DISTAL RIGHT CORONARY ARTERY (RCA). THE PHYSICIAN PREDILATED WITH A 3.0X12MM MAVERICK BALLOON. THE PHYSICIAN DEPLOYED A 3.5X12MM TAXUS EXPRESS2 DRUG ELUTING STENT, INFLATED TO 14 ATM, WITH NICE ANGIOGRAPHIC RESULTS. FOLLOW-UP ANGIOGRAPHY FOUND A SLIGHT ECCENTRICITY TO THE STENT WITH AN INDENTATION ALONG THE LESSER CURVATURE. POST DILATATION WAS PERFORMED WITH A 3.5X8MM QUANTUM MAVERICK BALLOON, INFLATED TO 15 ATM. THE PT EXPERIENCED CHEST PAIN FOLLOWING THE DEPLOYMENT OF THE STENT, WHICH WAS TREATED WITH INTRACORONARY VASODILATORS, VERAPAMIL, AND NITROGLYCERIN. GOOD ANGIOGRAPHIC RESULTS WERE ACHIEVED. MEDICATION GIVEN: HEPARIN AND ORAL PLAVIX. THE PT WAS DISCHARGED THE FOLLOWING DAY. THREE DAYS POST THE DEPLOYMENT OF THE STENT IN THE RCA, THE PT PRESENTED WITH CHEST PAIN. THE PT WAS GIVEN HEPARIN AND ADMITTED TO THE HOSPITAL. THE CHEST PAIN RESOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TAXUS EXPRESS2 PACLITAXEL-ELUTING CORONARY STENT SYSTEM | NIQ STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC | 3.50X12MM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |