FDA Adverse Event Injury Summary report: N

TAXUS EXPRESS2 PACLITAXEL-ELUTING CORONARY STENT SYSTEM

MDR report key: 1192122 · Received October 7, 2008

Report

Report Number
2134265-2008-02948
Event Type
Injury
Date Received
October 7, 2008
Date of Event
November 14, 2005
Report Date
September 23, 2008
Manufacturer
BOSTON SCIENTIFIC
Product Code
NIQ
PMA / PMN Number
P030025
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
DC, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINANT INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVAL; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFO FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Description of Event or Problem · 1

SAME CASE AS: 2134265-2008-02614. IT WAS REPORTED, THAT POST A CORONARY DRUG ELUTING STENTING TREATMENT PROCEDURE, CHEST PAIN AND HOSPITALIZATION OCCURRED. THE 70-80% STENOSED LESION BEING TREATED WAS LOCATED IN THE TORTUOUS, LIGHTLY CALCIFIED DISTAL RIGHT CORONARY ARTERY (RCA). THE PHYSICIAN PREDILATED WITH A 3.0X12MM MAVERICK BALLOON. THE PHYSICIAN DEPLOYED A 3.5X12MM TAXUS EXPRESS2 DRUG ELUTING STENT, INFLATED TO 14 ATM, WITH NICE ANGIOGRAPHIC RESULTS. FOLLOW-UP ANGIOGRAPHY FOUND A SLIGHT ECCENTRICITY TO THE STENT WITH AN INDENTATION ALONG THE LESSER CURVATURE. POST DILATATION WAS PERFORMED WITH A 3.5X8MM QUANTUM MAVERICK BALLOON, INFLATED TO 15 ATM. THE PT EXPERIENCED CHEST PAIN FOLLOWING THE DEPLOYMENT OF THE STENT, WHICH WAS TREATED WITH INTRACORONARY VASODILATORS, VERAPAMIL, AND NITROGLYCERIN. GOOD ANGIOGRAPHIC RESULTS WERE ACHIEVED. MEDICATION GIVEN: HEPARIN AND ORAL PLAVIX. THE PT WAS DISCHARGED THE FOLLOWING DAY. THREE DAYS POST THE DEPLOYMENT OF THE STENT IN THE RCA, THE PT PRESENTED WITH CHEST PAIN. THE PT WAS GIVEN HEPARIN AND ADMITTED TO THE HOSPITAL. THE CHEST PAIN RESOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TAXUS EXPRESS2 PACLITAXEL-ELUTING CORONARY STENT SYSTEM NIQ STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC 3.50X12MM

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization