LUPINE DRILL GUIDE
Report
- Report Number
- 1221934-2008-00467
- Event Type
- Malfunction
- Date Received
- October 1, 2008
- Date of Event
- September 29, 2008
- Report Date
- September 29, 2008
- Manufacturer
- DEPUY MITEK
- Product Code
- HRX
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
WE BELIEVE THAT THIS TYPE OF EVENT IS MOST LIKELY TECHNIQUE RELATED, EXTENSIVE LAB TESTING HAS SHOWN THAT UNDER NORMAL CONDITIONS THE REPORTED EVENT MODE COULD NOT BE DUPLICATED, HOWEVER, UNDER EXTREME CONDITIONS, THAT IS WHEN THE DRILL GUIDE WAS BENT DURING DRILLING WITH A DRILL BIT, THIS CAUSED THE DRILL BIT TO RUB AGAINST INNER SURFACE OF THE BENT DRILL GUIDE CAUSING SOME METAL TO SHAVE OFF OF THE DRILL GUIDE, THE REPORTED CONDITION WAS THEN DUPLICATED. THERE HAS BEEN SOME MFG PROCESSES CHANGES MADE TO SUBVERT AND OR GREATLY REDUCE THIS TYPE OF EVENT. THE PHENOMENA IS STILL UNDER STUDY BY THE MITEK QAE GROUP AND WHENEVER POSSIBLE RELATIVE FAILURE MODE DEVICES WILL BE FORWARDED TO THE GROUP TO SUBSIDIZE THEIR STUDY, ALSO, MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD. WHEN AND IF THE COMPLAINT DEVICE(S) IS RECEIVED HERE AT DEPUY MITEK IT WILL BE SUBJECTED TO A ROOT CAUSE FAILURE ANALYSIS. IF THE ANALYSIS IDENTIFIES ANY OTHER DEFINITIVE ROOT CAUSE A FOLLOW-UP REPORT WILL BE FILED.
OUR REP REPORTS THAT DURING AN ARTHROSCOPIC SHOULDER REPAIR, THE SURGEON NOTED THAT WITH THE USE OF THE LUPINE DRILL GUIDE, METAL SHAVINGS WERE FALLING INTO THE PT'S JOINT SPACE. THE SURGEON SWITCHED DRILL GUIDES, USING FOUR DIFFERENT ONES, HOWEVER, THE SAME PHENOMENA HAPPENED EACH TIME. ALL OF THE DEBRIS WAS REMOVED AND THE PROCEDURES WERE COMPLETED SUCCESSFULLY WITHOUT FURTHER ISSUE OR HARM TO THE PT. COMPLAINT DEVICES DISCARDED BY USER FACILITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LUPINE DRILL GUIDE | ARTHROSCOPIC INSTRUMENT | HRX | DEPUY MITEK | 213819 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |