FDA Adverse Event Malfunction Summary report: N

PROGRAMMING WAND

MDR report key: 1191906 · Received October 1, 2008

Report

Report Number
1644487-2008-02186
Event Type
Malfunction
Date Received
October 1, 2008
Date of Event
January 1, 2008
Report Date
September 5, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

NO PROBLEMS WERE NOTED PRIOR TO RETURN OF DEVICE, BUT WAND WAS RETURNED TO MFR AND UNDERWENT PRODUCT ANALYSIS. UPON ANALYSIS, IT WAS DISCOVERED THAT THE SERIAL DATA CABLE HAD OPEN AND INTERMITTENT CONDUCTORS WHICH CAUSED COMMUNICATION ERRORS. SERIAL DATA CABLE WAS REPLACED WITH KNOWN GOOD BENCH CABLE AND ALL COMMUNICATION ERRORS CLEARED. EVERYTHING ELSE WAS OPERATING WITHIN NORMAL LIMITS AND NO VISUAL ANOMALIES WERE OBSERVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING WAND LYJ CYBERONICS, INC. 201

Patients

Seq Age Sex Outcome Treatment
1