INPEN MMT-105NNGYNA NOVO NORDISK GRAY
Report
- Report Number
- 3012822846-2021-00361
- Event Type
- Malfunction
- Date Received
- June 1, 2021
- Date of Event
- February 11, 2019
- Report Date
- June 22, 2021
- Manufacturer
- COMPANION MEDICAL INC
- Product Code
- FMF
- UDI-DI
- 000010862088000351
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
SAND OR OTHER FOREIGN DEBRIS CAUSING EE HOUSING AND KNOB TO BIND. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. THIS MDR IS RELATED TO THE RETROSPECTIVE REVIEW OF COMPLAINTS FROM COMPANION MEDICAL INC, FOLLOWING MEDTRONIC¿S ACQUISITION OF THE COMPANY IN SEPTEMBER 2020.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WERE HAVING ISSUES WITH THE SCREW OF THE INSULIN PEN AND NO LONGER DELIVERING INSULIN. THE CUSTOMER STATED WHEN HE DIALED TO DESIRED UNITS OF INSULIN, THE SCREW WOULD RETRACT AND WOULD NOT DELIVER INSULIN. CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PEN WILL NOT BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 814180 | INPEN MMT-105NNGYNA NOVO NORDISK GRAY | SYRINGE, PISTON | FMF | COMPANION MEDICAL INC | MMT-105NNGYNA | A92FN | 000010862088000351 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |