SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL
Report
- Report Number
- 2023826-2008-01254
- Event Type
- Malfunction
- Date Received
- October 1, 2008
- Date of Event
- July 16, 2008
- Report Date
- September 10, 2008
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- HQL
- PMA / PMN Number
- P880091
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
EVAL CODES: RESULTS: OTHER - VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND THE LENS STUCK IN A RETURNED CARTRIDGE. THERE WAS NO VISIBLE DAMAGE TO THE CARTRIDGE. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. IT SHOULD BE NOTED THAT THE INJECTOR WAS NOT RETURNED FOR EVAL. CONCLUSIONS: AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN JUNE 2005. THE DAMAGE TO THE LENS, AS OBSERVED AND PHOTOGRAPHED UPON THE RETURN OF THE LENS TO STAAR, CANNOT BE DEFINITIVELY AND EXCLUSIVELY CORRELATED TO A SPECIFIC ROOT CAUSE. POSSIBLE ROOT CAUSES FOR THE TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED, AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.
THE REPORTER STATED THE SURGEON ATTEMPTED TO INSERT AN AA4204VF SILICONE SINGLE PIECE LENS BUT THE LENS BECAME STUCK IN THE CARTRIDGE. THERE WAS NO PT CONTACT OR INJURY. THE FACILITY WAS UNABLE TO PROVIDE ANY ADDITIONAL INFO. IF ADDITIONAL INFO BECOMES AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL | INTRAOCULAR LENS | HQL | STAAR SURGICAL CO. | AA4204VF | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CARTRIDGE MODEL MTC-60C FP - LOT NUMBER UNK| INJECTOR MODEL AND LOT NUMBER - UNK |