FDA Adverse Event Malfunction Summary report: N

3010188-020 QC RTS 1/P 10/B 50

MDR report key: 11913339 · Received June 1, 2021

Report

Report Number
1219103-2021-00323
Event Type
Malfunction
Date Received
June 1, 2021
Date of Event
May 26, 2021
Report Date
February 7, 2022
Manufacturer
COVIDIEN
Product Code
DRO
UDI-DI
00721902195037
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HISTORY RECORDS (DHR) WERE REVIEWED FOR BOTH THE FINISHED GOOD AND THE SUB-ASSEMBLIES MANUFACTURED AT THE PLANT, AND NO ABNORMAL PROCESS CONDITIONS WERE PRESENT DURING THE MANUFACTURING OF THE PRODUCT THAT COULD HAVE LED TO THE REPORTED CONDITION AS DESCRIBED BY THE CUSTOMER. THE CUSTOMER PROVIDED A PICTURE OF THE ISSUE WHERE A SMALL PORTION OF THE GEL DELAMINATED FROM THE SILVER AREA OF THE GEL BODY. AN ADDITIONAL PICTURE OF THE LOT NUMBER LABEL ON THE POUCH INDICATED THAT THE GEL BODY USED WOULD HAVE COME FROM LOT 934354. SAMPLES WERE RECEIVED FROM THE CUSTOMER IN THE FORM OF 9 UNOPENED/SEALED POUCHES. A VISUAL INSPECTION OF THE POUCHES DID NOT SHOW ANY DAMAGE TO THE POUCH OR THE SEAL. ADDITIONALLY, A VISUAL INSPECTION OF THE PRODUCT PRIOR TO TESTING DID NOT SHOW ANY SIGNS OF DAMAGE OR ANYTHING THAT MAY HAVE INDICATED AN ISSUE WITH THE PRODUCT. A TRACEABILITY NUMBER ON THE BACK OF THE POUCHES INDICATED THAT THE RETURNED SAMPLES WOULD HAVE BEEN MADE FROM GEL BODIES FROM LOT 934354. PHYSICAL TESTING WAS PERFORMED ON THE SAMPLES USING THE SINGLE DIRECTION PEEL TEST. THIS IS A TEST THAT USES AN INSTRON TO PEEL THE GEL BODY FROM THE LINER AND TO INSPECT FOR GEL DELAMINATION. THIS TEST RESULTED IN DELAMINATION ON THE SAMPLES RETURNED, WITH OVER 50 PERCENT OF THE GEL DELAMINATING FROM THE SILVER PORTION OF THE GEL BODY. THE TEST RESULTS FROM THE NINE RETURNED SAMPLES SHOWED DELAMINATION WHERE THE GEL HAD A TIGHTER BOND TO THE LINER THAN THE SILVER AND CARBON VINYL THAT IT IS DESIGNED TO BE TIGHTLY BONDED TO. IT IS IMPORTANT TO NOTE THAT GEL DELAMINATION IS ADDRESSED IN ALL RISK DOCUMENTATION ASSOCIATED WITH DEFIBRILLATION ELECTRODES. THE RESULTS OF THE MANUFACTURING FACILITY INVESTIGATION WERE UNABLE TO CONFIRM ANY POTENTIAL ROOT CAUSES ASSOCIATED WITH THE MANUFACTURE OF PRODUCT WHICH WOULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. INDICATIONS FROM THE RETURNED SAMPLES INDICATE A LOCALIZED PROBLEM DURING THE GEL BODY MANUFACTURING PROCESS. HOWEVER, BECAUSE IT WAS IN THE RETURNED SAMPLES BUT NOT IN THE RETAINS FROM THE SAME LOT IT WAS NOT POSSIBLE TO VERIFY WITH CERTAINTY. NO CORRECTIVE OR PREVENTATIVE ACTIONS WILL BE TAKEN AT THIS TIME. WE WILL CONTINUE TO TREND THIS ISSUE FOR FUTURE OCCURRENCES AS PART OF THE COMPLAINT REVIEW PROCESS.

Additional Manufacturer Narrative · 1

THE INCIDENT SAMPLE HAS BEEN REQUESTED BUT TO DATE HAS NOT BEEN RECEIVED FOR EVALUATION.  IF THE SAMPLE IS RECEIVED, OR IF ADDITIONAL INFORMATION PERTINENT TO THE INCIDENT IS OBTAINED A FOLLOW-UP REPORT WILL BE SUBMITTED.  AS PART OF OUR MANUFACTURING PROCESS, ALL DEVICE HISTORY RECORDS ARE REVIEWED AND APPROVED BY QUALITY, PRIOR TO RELEASE OF PRODUCT.

Description of Event or Problem · 1

CUSTOMER REPORTS THAT ONE OF THEIR CUSTOMERS REPORTED THAT AN ELECTRODE WAS FOUND TO BE EXPERIENCING ADHESIVE/GEL DAMAGE WHEN THE PAD WAS PEELED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
813931 3010188-020 QC RTS 1/P 10/B 50 PACEMAKER, CARDIAC, EXTERNAL TRANSCUTANEOUS (NON-INVASIVE) DRO COVIDIEN PM20003 004827 00721902195037

Patients

Seq Age Sex Outcome Treatment
1 Unknown