HARMONIC SCALPEL GENERATOR
Report
- Report Number
- 1527736-2008-03642
- Event Type
- Malfunction
- Date Received
- October 9, 2008
- Date of Event
- May 19, 2008
- Report Date
- May 22, 2008
- Manufacturer
- ETHICON ENDO SURGERY, INC. (CINCINNATI)
- Product Code
- LFL
- PMA / PMN Number
- K002906
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
EVAL SUMMARY: BASED ON THE ANALYSIS RESULTS, THIS COMPLAINT IS NOW DETERMINED TO BE A MDR MALFUNCTION. THE ANALYSIS SITE CONFIRMED THE CUSTOMER COMPLAINT AND FOUND THE MAIN PCB WAS CAUSING THE UNIT TO LOSE DISPLAY INTERMITTENTLY. TO CORRECT THIS ISSUE, THE ANALYSIS SITE REPLACED THE MAIN PCB. THE SITE ALSO FOUND THE DISPLAY ON THE BEZEL SUB-ASSEMBLY WAS DAMAGED AND TO CORRECT THIS ISSUE, THE ANALYSIS SITE REPLACED THE BEZEL SUB-ASSEMBLY. PER THE SERVICE MANUAL, CALIBRATION AND TEST WERE PERFORMED WITH THE UNIT PASSING ALL TESTS. THE DEVICE HISTORY RECORD HAS BEEN REVIEWED AND HAS PASSED QA INSPECTION. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED.
IT WAS REPORTED THAT THE DISPLAY ON THE GENERATOR WOULD DISAPPEAR WHEN SOMEONE TAPPED ON THE DISPLAY. THE GENERATOR WAS SWAPPED PRIOR TO THE START OF AN UNK CASE AND THE CASE WAS COMPLETED WITH NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC SCALPEL GENERATOR | NONE | LFL | ETHICON ENDO SURGERY, INC. (CINCINNATI) | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HANDPIECE |