FDA Adverse Event Malfunction Summary report: N

HARMONIC ACE 36 CM

MDR report key: 1191207 · Received October 9, 2008

Report

Report Number
1527736-2008-03631
Event Type
Malfunction
Date Received
October 9, 2008
Date of Event
February 13, 2008
Report Date
May 14, 2008
Manufacturer
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DATE SENT: 06/10/2008. THE DEVICE WAS RECEIVED IN GOOD CONDITION. NO VISIBLE DAMAGE WAS NOTED. THE HAND-ACTIVATION CONTACTS WERE IN GOOD CONDITION. THE DEVICE WAS TESTED ON A GENERATOR AND NO SOLID TONE WAS NOTED; HOWEVER, AN ERROR CODE 7 WAS DISPLAYED. IT WAS FOUND THAT THE HAND CONTROL SWITCH ASSEMBLY BUTTONS WERE NOT FUNCTIONAL. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A GYN PROCEDURE, THE DEVICE KEPT ALARMING. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO CONSEQUENCE TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HARMONIC ACE 36 CM NONE LFL ETHICON ENDO SURGERY, INC. (CINCINNATI) NA D4JZ79

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR| HANDPIECE