FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC SCALPEL HAND PIECE

MDR report key: 1191009 · Received October 9, 2008

Report

Report Number
1527736-2008-03568
Event Type
Malfunction
Date Received
October 9, 2008
Report Date
April 24, 2008
Manufacturer
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Product Code
LFL
PMA / PMN Number
K002906
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL SUMMARY: THE HANDPIECE WAS RETURNED WITH A LOOSE MOUNT AND THE NOSE CONE CRACKED. IT WAS TESTED ON A GEN04 AND GAVE ERROR #5; NO FURTHER EVAL WITH THE GEN04 COULD BE PERFORMED. THE HANDPIECE WAS DISASSEMBLED, A TORN ACOUSTIC ISOLATOR AND EVIDENCE OF MOISTURE INGRESS WERE FOUND IN THE INSTRUMENT. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MFG PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THE DEVICE BECAME HOT. UNK HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC SCALPEL HAND PIECE NONE LFL ETHICON ENDO SURGERY, INC. (CINCINNATI) NA NA

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR