FDA Adverse Event Malfunction Summary report: N

HARMONIC ACE 36 CM

MDR report key: 1190762 · Received October 9, 2008

Report

Report Number
1527736-2008-03486
Event Type
Malfunction
Date Received
October 9, 2008
Report Date
April 18, 2008
Manufacturer
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE A WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKE OFF AND PRESENT. THE REMAINING BLADE PORTION WAS SCRATCHED. THIS BLADE TIP PORTION MAY HAVE BROKEN OFF OF THE DEVICE DURING TRANSPORT TO OUR DECONTAMINATION FACILITY OR FROM OUR DECONTAMINATION TO THE ANALYSIS SITE. THE REPORTED COMPLAINT WAS "ACTIVATION ISSUES". THE DEVICE WILL STOP ACTIVATING, AND EITHER EMIT A SOLID TONE OR DISPLAY AN INSTRUMENT ERROR CODE 5 ON THE GENERATOR DISPLAY WHEN THE BLADE BECOMES DAMAGED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. THEREFORE, THE INSTRUCTIONAL INSERT STATES: "AVOID ACCIDENTAL CONTACT WITH ALL METAL OR PLASTIC INSTRUMENTS OR OBJECTS WHEN THE INSTRUMENT IS ACTIVATED. CONTACT WITH STAPLES, CLIPS OR OTHER INSTRUMENTS WHILE THE INSTRUMENT IS ACTIVATED AMY RESULT IN CRACKED OR BROKEN BLADES, WHICH MAY BE IDENTIFIED BY GENERATOR SOLID TONE OR INSTRUMENT ERROR." THE ANALYSIS RESULTS FOUND THAT THE DEVICE B WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. THE DEVICE WILL STOP ACTIVATING, AND EITHER EMIT A SOLID TONE OR DISPLAY AN INSTRUMENT ERROR CODE 5 ON THE GENERATOR DISPLAY WHEN THE BLADE BECOMES DAMAGED. DUE TO THE DAMAGE TO THE BLADE, IT WAS CONFIRMED THAT THE DEVICE WAS NON-FUNCTIONAL. THE DEVICE WAS TESTED WITH A GEN04 AND AN ERROR CODE 5 WAS NOTED. WHEN A BLADE HAS BEEN COMPROMISED, SCRATCHED OR CRACKED, THE TITANIUM METAL IS FATIGUED WHEN CONTINUALLY ENERGIZED AND THIS RESULTS IN THE BLADE FURTHER CRACKING AND POSSIBLY BREAKING OFF. DURING THE ANALYSIS PROCESS, THE BLADE WAS TESTED FOR SCRATCHES AND CRACKS AND THE BLADE BROKE OFF. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. FOR THIS REASON, THE FOLLOWING STATEMENTS WERE INCLUDED IN THE INSTRUCTIONS FOR USE: "AVOID ACCIDENTAL CONTACT WITH ALL METAL OR PLASTIC INSTRUMENTS OR OBJECTS WHEN THE INSTRUMENT IS ACTIVATED. CONTACT WITH STAPLES, CLIPS OR OTHER INSTRUMENTS WHILE THE INSTRUMENT IS ACTIVATED MAY RESULT IN CRACKED OR BROKEN BLADES, WHICH MAY BE IDENTIFIED BY GENERATOR SOLID TONE OR INSTRUMENT ERROR." EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED AT THIS PROCESS. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MFG PROCESS. DEVICE B BATCH E9EL1N. MFG DATE 01/2008. EXP DATE 12/2012.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAP SIGMA PROCEDURE, THE DEVICE DID NOT FUNCTION. ANOTHER LIKE DEVICE WAS USED. THERE WAS NO PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HARMONIC ACE 36 CM NONE LFL ETHICON ENDO SURGERY, INC. (CINCINNATI) NA D4JG67

Patients

Seq Age Sex Outcome Treatment
1 HANDPIECE| GENERATOR