HARMONIC FOCUS CURVED SHEARS, TORQUE WRENCH AND GRIP ASSIST
Report
- Report Number
- 3005075853-2008-02157
- Event Type
- Malfunction
- Date Received
- October 7, 2008
- Report Date
- September 15, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- LFL
- PMA / PMN Number
- K063192
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD CONDITION. THE DEVICE WAS TESTED WITH A GENERATOR AND WAS FUNCTIONAL. THE TISSUE PAD WAS EXAMINED AND NORMAL WEAR WAS VISUALLY SEEN. FLAKING OF THE TISSUE PAD MAY BE CAUSED DUE TO ACTIVATION OF THE DEVICE WHILE CLAMPING WITHOUT TISSUE BEING PRESENT IN THE ENTIRE JAW AREA. CARE SHOULD BE TAKEN NOT TO APPLY PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. THIS CAN RESULT IN DAMAGE TO THE INSTRUMENT. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A THYROIDECTOMY PROCEDURE, THE DEVICES TISSUE PAD WAS MELTING WHILE TISSUE WAS IN THE JAWS. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC FOCUS CURVED SHEARS, TORQUE WRENCH AND GRIP ASSIST | LFL | ETHICON ENDO-SURGERY, LLC | NA | E4KN83 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HANDPIECE| GENERATOR |