FDA Adverse Event
Malfunction
Summary report: N
ESPRIT VENTILATOR
MDR report key: 1190624
·
Received October 6, 2008
Report
- Report Number
- 2031642-2008-00207
- Event Type
- Malfunction
- Date Received
- October 6, 2008
- Date of Event
- September 8, 2008
- Report Date
- September 8, 2008
- Manufacturer
- RESPIRONICS CALIFORNIA, INC.
- Product Code
- CBK
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EXHALATION FLOW SENSOR.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE VENTILATOR WENT VENT INOP AND ALARMED WHILE IN USE ON A PATIENT. THE CUSTOMER REPORTED THERE WAS NO PATIENT HARM. VENT INOP, WHEN IN USE, WILL STOP PROVIDING VENTILATORY SUPPORT TO THE PATIENT. THE MANUFACTURER'S SERVICE TECHNICIAN WAS UNABLE TO DUPLICATE THE CUSTOMER REPORTED PROBLEM; HOWEVER, THE SERVICE TECHNICIAN CONFIRMED A DIAGNOSTIC CODE IN LOG HISTORY THAT INDICATED A VENT INOP OCCURRED DUE TO A FLOW SENSOR FAILURE. THE SERVICE TECHNICIAN REPLACED THE EXHALATION FLOW SENSOR AS A PRECAUTION. EXTENDED SELF TESTING (EST) AND PERFORMANCE VERIFICATION TESTING WAS COMPLETED AND ALL TESTS PASSED PER OPERATING SPECIFICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ESPRIT VENTILATOR | VENTILATOR, CONTINUOUS | CBK | RESPIRONICS CALIFORNIA, INC. | V1000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | HME| DISPOSABLE PATIENT CIRCUIT| SPU EXPIRATORY FILTER |