FDA Adverse Event Malfunction Summary report: N

ESPRIT VENTILATOR

MDR report key: 1190624 · Received October 6, 2008

Report

Report Number
2031642-2008-00207
Event Type
Malfunction
Date Received
October 6, 2008
Date of Event
September 8, 2008
Report Date
September 8, 2008
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EXHALATION FLOW SENSOR.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE VENTILATOR WENT VENT INOP AND ALARMED WHILE IN USE ON A PATIENT. THE CUSTOMER REPORTED THERE WAS NO PATIENT HARM. VENT INOP, WHEN IN USE, WILL STOP PROVIDING VENTILATORY SUPPORT TO THE PATIENT. THE MANUFACTURER'S SERVICE TECHNICIAN WAS UNABLE TO DUPLICATE THE CUSTOMER REPORTED PROBLEM; HOWEVER, THE SERVICE TECHNICIAN CONFIRMED A DIAGNOSTIC CODE IN LOG HISTORY THAT INDICATED A VENT INOP OCCURRED DUE TO A FLOW SENSOR FAILURE. THE SERVICE TECHNICIAN REPLACED THE EXHALATION FLOW SENSOR AS A PRECAUTION. EXTENDED SELF TESTING (EST) AND PERFORMANCE VERIFICATION TESTING WAS COMPLETED AND ALL TESTS PASSED PER OPERATING SPECIFICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ESPRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK HME| DISPOSABLE PATIENT CIRCUIT| SPU EXPIRATORY FILTER