FDA Adverse Event Malfunction Summary report: N

LIFEPAK CR PLUS DEFIBRILLATOR

MDR report key: 1190623 · Received October 3, 2008

Report

Report Number
3015876-2008-01454
Event Type
Malfunction
Date Received
October 3, 2008
Date of Event
September 4, 2008
Report Date
September 4, 2008
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K011144
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE ANALOG PCB ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. PHYSIO-CONTROL FURTHER EVALUATED THE REPLACED ASSEMBLY AND DETERMINED THAT THE ROOT CAUSE OF THE REPORTED FAILURE WAS A LEAKY CAPACITOR, DESIGNATOR C177. THE CUSTOMER RECEIVED A REPLACEMENT DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD A LOW BATTERY CAPACITY AND THE CHARGE PACK AND ATTENTION ICONS WERE ILLUMINATED. THERE WAS NO PT USE ASSOCIATED WITH THIS EVENT.