FDA Adverse Event Malfunction Summary report: N

LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES

MDR report key: 1190612 · Received October 3, 2008

Report

Report Number
3015876-2008-01442
Event Type
Malfunction
Date Received
October 3, 2008
Date of Event
September 3, 2008
Report Date
September 3, 2008
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K991910
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE WAS VERIFIED THE REPORTED INTERMITTENT FAILURE. PHYSIO REPLACED THE SYSTEM AND THE MEMORY PCB ASSEMBLIES. PROPER DEVICE OPERATION WAS THEN OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO-CONTROL FURTHER EVALUATED THE REMOVED ASSEMBLIES AND OBSERVED THAT BOTH ASSEMBLIES WERE FUNCTIONING PROPERLY. THE REPORTED CONDITION COULD NOT BE DUPLICATED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD NOT POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES MKJ PHYSIO-CONTROL, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA