FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK 500 AUTOMATED EXTERNAL DEFIBRILLATOR
MDR report key: 1190597
·
Received October 3, 2008
Report
- Report Number
- 3015876-2008-01446
- Event Type
- Malfunction
- Date Received
- October 3, 2008
- Date of Event
- September 4, 2008
- Report Date
- September 4, 2008
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- MKJ
- PMA / PMN Number
- K983393
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE MAIN PC ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO-CONTROL FURTHER EVALUATED THE REPLACED ASSEMBLY AND DETERMINED THAT THE ROOT CAUSE OF THE REPORTED FAILURE WAS AN IC CHIP, DESIGNATOR U8, THAT WAS CAUSING THE ASSEMBLY TO BOOT UP TO A BLUE BAR AND THEN LOCK UP IN THAT STATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD NOT POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 500 AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | PHYSIO-CONTROL, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |