FDA Adverse Event Malfunction Summary report: N

BIPOLAR LEAD MODEL 300

MDR report key: 1190592 · Received October 6, 2008

Report

Report Number
1644487-2008-02418
Event Type
Malfunction
Date Received
October 6, 2008
Date of Event
September 1, 2008
Report Date
September 8, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ONLY A PORTION OF THE LEAD WAS RETURNED FOR ANALYSIS WHICH DID NOT REVEAL ANY ANOMALIES. DEVICE FAILURE IS SUSPECTED IN THE LEAD PORTIONS NOT RETURNED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

IT WAS REPORTED THE PT HAD THEIR VNS GENERATOR REPLACED FOR END OF BATTERY LIFE AND THEIR LEAD REPLACED FOR LEAD BREAK. THE GENERATOR WAS RETURNED FOR ANALYSIS AND MET MFR SPECS. THE LEAD WAS RETURNED IN THREE PORTIONS. ANALYSIS WAS PERFORMED ON THE RETURNED LEAD PORTIONS AND THE REPORTED "FRACTURE OF LEAD" CONDITION WAS NOT VERIFIED. A LARGE PORTION OF THE LEAD ASSEMBLY BODY INCLUDING THE ELECTRODES WAS NOT RETURNED FOR ANALYSIS; THEREFORE, A COMPLETE EVAL COULD NOT BE PERFORMED ON THE ENTIRE LEAD PROD. ANALYSIS WAS PERFORMED ON THE RETURNED LEAD PORTIONS. WHAT APPEARED TO BE WHITE DEPOSITS WERE OBSERVED ON THE MARKED AND UNMARKED CONNECTOR BOOTS, CONNECTOR SILICONE TUBES AND OUTER SILICONE TUBING. THE CONDITION OF THE RETURNED LEAD PORTIONS IS CONSISTENT WITH CONDITIONS THAT TYPICALLY EXIST FOLLOWING AN EXPLANT PROCEDURE. NO OBVIOUS ANOMALIES WERE NOTED. THE RESISTANCE MEASUREMENT TAKEN DURING DECONTAMINATION VERIFIED AN ELECTRICAL AND MECHANICAL CONTACT BETWEEN THE GENERATOR AND CONNECTOR PINS AT ONE POINT IN TIME. CONTINUITY CHECKS OF THE RETURNED LEAD PORTIONS WERE PERFORMED, DURING THE VISUAL ANALYSIS, AND NO DISCONTINUITIES WERE IDENTIFIED. DEVICE FAILURE IS SUSPECTED ON THE PORTION OF THE LEAD NOT RETURNED. GOOD FAITH ATTEMPTS ARE BEING MADE FOR ADD'L DETAILS SURROUNDING THE REPORTED LEAD BREAK. THUS FAR NO FURTHER INFO HAS BEEN RECEIVED FROM THE SITE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BIPOLAR LEAD MODEL 300 LYJ CYBERONICS, INC. 300-20 13389C

Patients

Seq Age Sex Outcome Treatment
1 23 YR