FDA Adverse Event
Malfunction
Summary report: N
AFFINITY BED
MDR report key: 1190544
·
Received October 2, 2008
Report
- Report Number
- 1824206-2008-03657
- Event Type
- Malfunction
- Date Received
- October 2, 2008
- Date of Event
- September 8, 2008
- Report Date
- September 8, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- HDD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN INDICATED THE BRAKES WILL NOT HOLD ON THE BED. THE TECHNICIAN ADJUSTED THE CASTERS TO REPAIR THE BED. CHAPTER 6 OF THE SERVICE MANUAL DOCUMENTS A FUNCTION CHECK FOR THE CASTER TIRES AND CENTRAL BRAKE AND STEER AS PART OF PREVENTATIVE MAINTENANCE. THE COMPONENTS SHOULD BE CHECKED FOR CUTS, WEAR, TREAD LIFE, ETC. AND THE BRAKES SHOULD BE CHECKED TO ENSURE THE BED WILL NOT MOVE WHEN THE BRAKE IS ACTIVATED. IF THE BED MOVES, INSPECTION FOR WEAR AND ADJUSTMENT MAY BE NECESSARY. THE STEERING PEDAL SHOULD ALSO BE CHECKED FOR PROPER LOCKING WHEN ACTIVATED.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THE BRAKES WILL NOT HOLD ON THE BED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AFFINITY BED | BIRTHING BED | HDD | HILL-ROM RITTER | 3600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |