FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 1190503 · Received October 9, 2008

Report

Report Number
1527736-2008-03430
Event Type
Malfunction
Date Received
October 9, 2008
Date of Event
December 5, 2007
Report Date
January 25, 2008
Manufacturer
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Product Code
GDO
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DATE SENT: 06/03/2008. THE ANALYSIS RESULTS FOUND THAT THE EL5ML DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A "J" CLIP ON THE JAWS, THE ROOT CAUSE OF A J-SHAPED MALFORMED CLIP IS FEEDING THE CLIP INTO A CLOSED SET OF JAWS. THE FOLLOWING ARE SCENARIOS OF HOW THIS MALFORMED CLIP COULD HAVE HAPPENED: HAVING THE JAWS CLOSED IN A TROCAR AND ACTUATING THE TRIGGER. HAVING THE JAWS CLOSED IN THE MOLDED ENDCAP AND ACTUATING THE TRIGGER. HAVING THE JAWS CLOSED BY BURYING TIPS IN TISSUE AND ACTUATING THE TRIGGER. IT SHOULD BE NOTED THAT THE JAWS NEED TO BE FULLY OPEN IN ORDER FOR THE NEXT CLIP TO BE PROPERLY FED. THE INSTRUMENT WAS CYCLED, FED AND FORMED THE REMAINING CLIPS WITHIN MFG SPECIFICATIONS. THE INSTRUMENT LOCKED OUT AS INTENDED AND IT OPENED AND CLOSED WITHOUT ANY DIFFICULTIES. HOWEVER, THE JAWS WERE NOTED TO BE OUT OF ACCEPTABLE LIMITS, THE DEVICE WAS DISSASSEMBLED AND NO ANOMALIES WERE FOUND. WHILE WE WERE UNABLE TO RE-CREATE THE EVENT, WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. NO INCIDENT RELATED TO THE REPORTED EVENT WAS OBSERVED DURING THE BATCH RECORD REVIEW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS USED DURING A LAPAROSCOPIC CHOLECYSTECTOMY. THE DEVICE FAILED. UNK HOW CASE COMPLETED. THERE WAS NO CONSEQUENCE TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM ENDO CLIP APPLIER NONE GDO ETHICON ENDO SURGERY, INC. (CINCINNATI) NA UNK

Patients

Seq Age Sex Outcome Treatment
1