MAMMOTOME CONTROL MODULE WITH SMARTVAC - 120V

MDR report key: 1190484 · Received October 9, 2008

Report

Report Number
1527736-2008-03416
Event Type
Malfunction
Date Received
October 9, 2008
Date of Event
May 19, 2008
Report Date
May 19, 2008
Manufacturer
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Product Code
KNW
PMA / PMN Number
K991980
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DATE SENT: 06/03/2008. INFO NOT PROVIDED. EVAL SUMMARY: BASED ON THE ANALYSIS RESULTS, THIS COMPLAINT IS NOW DETERMINED TO BE A MDR MALFUNCTION. THE ANALYSIS SITE CONFIRMED THE CUSTOMER'S COMPLAINT AND FOUND THE DC ADAPTER WAS THE CAUSE. TO CORRECT THE CUSTOMER'S COMPLAINT THE ANALYSIS SITE REPLACED THE DC ADAPTER. PER THE SERVICE MANUAL PERFORMED SERVICE TESTS, WITH NO FUNCTIONAL FAULTS FOUND. AS PART OF THE STANDARD SERVICE PROCESS, REPLACED THE MUFFLER AND APPLIED DOW CORNING LUBRICANT TO THE DC MOTORS. THE UNIT HAS NOT BEEN RETURNED FOR SERVICE PRIOR TO THIS EVENT, THEREFORE, NO SERVICE HISTORY REVIEW CAN BE DONE. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A BREAST BIOPSY PROCEDURE VERY SMALL FRAGMENTED SAMPLES WERE BEING RECEIVED. THERE WAS NO PT CONSEQUENCE REPORTED. CASE WAS COMPLETED USING THE UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAMMOTOME CONTROL MODULE WITH SMARTVAC - 120V NONE KNW ETHICON ENDO SURGERY, INC. (CINCINNATI) NA NA

Patients

Seq Age Sex Outcome Treatment
1 PROBE| HOLSTER