FDA Adverse Event Injury Summary report: N

PINNACLE SECTOR II CUP 56MM

MDR report key: 1190382 · Received October 3, 2008

Report

Report Number
1818910-2008-04292
Event Type
Injury
Date Received
October 3, 2008
Date of Event
September 3, 2008
Report Date
March 10, 2017
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
KWA
PMA / PMN Number
K073504
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EXAMINATION WAS NOT POSSIBLE, AS THE DEVICE WAS NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT CODE SINCE ITS RELEASE FOR DISTRIBUTION. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT ERROR WITH REGARD TO THE REPORTED EVENT. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THIS INVESTIGATION CLOSED. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION THAT CHANGES THIS CONCLUSION BE RECEIVED, THE INVESTIGATION WILL BE REOPENED.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # PC-(B)(4). INVESTIGATION SUMMARY NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND NO OTHER REPORTED INCIDENT(S) AGAINST THE PROVIDED PRODUCT/LOT COMBINATION(S) SINCE RELEASE FOR DISTRIBUTION. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND NO OTHER REPORTED INCIDENT(S) AGAINST THE PROVIDED PRODUCT/LOT COMBINATION(S) SINCE RELEASE FOR DISTRIBUTION. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

PATIENT REVISED TO ADDRESS MALPOSITIONED CUP AND SQUEAKING.

Description of Event or Problem · 1

UPDATE: 03/10/2017 - PFS AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF THE PFS AND MEDICAL RECORDS FOR MDR REPORTABILITY, THE PATIENT WAS REVISED TO ADDRESS PAIN, DISCOMFORT, AND RECURRENT DISLOCATING RIGHT HIP. THE FEMORAL HEAD HAD SCRATCHING FROM IMPINGEMENT AND METAL DEBRIS WAS REPORTED. METAL ION LEVELS PROVIDED WERE AT NON-REPORTABLE LEVELS. THE COMPLAINT WAS UPDATED: 03/31/2017.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PINNACLE SECTOR II CUP 56MM 87KWA KWA DEPUY ORTHOPAEDICS, INC. 1818910 N/A XT1EW1002

Patients

Seq Age Sex Outcome Treatment
1 51 YR Required Intervention