FDA Adverse Event Injury Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 11903119 · Received May 28, 2021

Report

Report Number
1221359-2021-01583
Event Type
Injury
Date Received
May 28, 2021
Date of Event
April 26, 2021
Report Date
August 12, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
10811877011320
PMA / PMN Number
EUA2000074
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO REPORT THE INVESTIGATION CONCLUSION AND ADDITIONAL INFORMATION. PLEASE UPDATES TO SECTION: G3, G6, H2 AND H6. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1022562 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 191-000 / LOT 1022562 TEST BASE PART NUMBER 190-430 / LOT 1022562. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVES RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1022562 SHOWED THAT THE COMPLAINT RATE IS (B)(4)%. IN CONCLUSION, ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION.

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. PLEASE REFER TO MFR. REPORT NUMBERS: 1221359-2021-01582, 1221359-2021-01584 AND 1221359-2021-01585.

Description of Event or Problem · 1

THE CUSTOMER REPORTED FOUR (4) FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY MULTIPLE PATIENTS PERFORMED ON (B)(6) 2021, (B)(6) 2021, (B)(6) 2021 AND (B)(6) 2021 USING TWO DIFFERENT LOTS. THIS MFR. REPORT ADDRESSES PATIENT 2 OF 4. THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021 ON A DIRECT TESTED ICLEAN NASOPHARYNGEAL SWAB. 4PLEX CEPHEID PCR CONFIRMATION TESTING ON (B)(6) 2021 USING A NASOPHARYNGEAL SWABS IN TRANSPORT MEDIA (M4RT) GENERATED NEGATIVE RESULTS. THE CUSTOMER STATED THE PATIENT WAS ASYMPTOMATIC . THE PATIENT WAS DELAYED AT HOSPITAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
797432 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1022562 10811877011320

Patients

Seq Age Sex Outcome Treatment
1 88 YR Hospitalization