Narratives

Description of Event or Problem · 1

PT DEVELOPED LEAK IN ETT CUFF. STAFF UNABLE TO INJECT AIR THROUGH PILOT BALLOON. PT EXTUBATED AND REINTUBATED. PILOT BALLOON TUBE FOUND KINKED UNDER BITE BLOCK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BITE-BLOCK BRW SONTEK MEDICAL, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 43 YR Required Intervention