FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP

MDR report key: 1190022 · Received October 8, 2008

Report

Report Number
6000001-2007-02878
Event Type
Malfunction
Date Received
October 8, 2008
Date of Event
January 1, 2007
Report Date
January 4, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA JAN 25, 2007. EVALUATION SUMMARY: FAILURE CODE 814:04 WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THAT CAUSE OF THE REPORTED FAILURE CODE WAS DUE TO A DEFECTIVE AIR-IN-LINE PRINTED CIRCUIT BOARD. THE AIR-IN-LINE PRINTED CIRCUIT BOARD WAS REPLACED. THE PUMP WAS RETURNED TO THE CUSTOMER FULLY OPERATIONAL. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

A BAXTER REPRESENTATIVE REPORTED TO CUSTOMER SERVICE A PUMP WITH FAILURE CODE 814:04. THIS FAILURE CODE OCCURRED DURING BIO-MED TESTING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1