INOMAX DSIR PLUS (DELIVERY SYSTEM)
Report
- Report Number
- 3004531588-2021-00136
- Event Type
- Malfunction
- Date Received
- May 28, 2021
- Date of Event
- January 30, 2021
- Report Date
- May 28, 2021
- Manufacturer
- MALLINCKRODT MANUFACTURING LLC.
- Product Code
- MRN
- UDI-DI
- 00867538000106
- PMA / PMN Number
- K061901
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- 003
Narratives
REPORTABLE MALFUNCTION WITH INOMAX DSIR PLUS (B)(4) WAS DOCUMENTED AND INVESTIGATED UNDER (B)(4). A REVIEW OF THE DEVICE'S SERVICE LOG REVEALED THAT A DELIVERY STOPPED ALARM OCCURRED AS THE MONITORED NITRIC OXIDE (NO) VALUE (I.E., ACTUAL VALUE: 119 PPM) WAS GREATER THAN THE ABSOLUTE MAX OF 100 PPM FOR 12 CONSECUTIVE SECONDS. AFTERWARDS, A SERVICE LOG ENTRY FOR A FAILED LOW NO CELL CALIBRATION DUE TO THE LOW POINTS BEING GREATER THAN THE MAXIMUM VALUE (MAX: 655 COUNTS; ACTUAL VALUE: 1358 COUNTS) WAS OBSERVED. ALTHOUGH IDENTIFIED IN THE SERVICE LOG, THE REGIONAL SERVICE CENTER (RSC) DID NOT EXPERIENCE A DELIVERY STOPPED ALARM DURING TESTING. THE ROOT CAUSE FOR THIS OCCURRENCE WAS LIKELY DUE TO A MALFUNCTIONING NO SENSOR. AS A RESULT, THE DEVICE'S NO SENSOR WAS REPLACED. TRENDS WERE REVIEWED FOR THIS REPORTABLE CONDITION AND DETERMINED TO BE WITHIN THE ESTABLISHED CONTROL LIMITS. NO FURTHER ACTION IS REQUIRED AT THIS TIME. A FULL FUNCTIONAL TEST WAS PERFORMED AND THE DEVICE OPERATED ACCORDING TO SPECIFICATIONS. AS A RESULT, THE DEVICE WAS PLACED BACK INTO CIRCULATION FOR CUSTOMER USE.
DURING A ROUTINE SERVICE LOG REVIEW FOR A DEVICE LOCATED IN THE UNITED STATES, A MALLINCKRODT EMPLOYEE DISCOVERED THAT A DELIVERY STOPPED ALARM HAD OCCURRED DUE TO A NITRIC OXIDE (NO) CONCENTRATION GREATER THAN 100 PPM FOLLOWED BY A FAILED LOW NO SENSOR CALIBRATION. THIS SERVICE LOG FINDING MEETS THE CRITERIA OF A REPORTABLE MALFUNCTION. THERE WAS NO DELIVERY STOPPED ALARM COMPLAINT OR REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 796718 | INOMAX DSIR PLUS (DELIVERY SYSTEM) | APPARATUS | MRN | MALLINCKRODT MANUFACTURING LLC. | 10007 | 00867538000106 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |