THERMAGE CPT SYSTEM
Report
- Report Number
- 3011423170-2021-00055
- Event Type
- Malfunction
- Date Received
- May 27, 2021
- Date of Event
- May 6, 2021
- Report Date
- May 6, 2021
- Manufacturer
- SOLTA MEDICAL, INC
- Product Code
- GEI
- PMA / PMN Number
- K132431
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE PRODUCT WAS RETURNED AND EVALUATED. BURNT TRACE WAS CONFIRMED ON THE TIP SURFACE. THIS IS A VALID TIP VERIFIED AGAINST THE SOLTA DATABASE. THE TIP START TIME WAS 4:0:40 PM ON 04/28/21 AND WAS USED FOR 247 TREATMENTS. THE TIP PASSED THE FLOW TEST AND FAILED THE LEAK TEST. THE TIP FAILED THE VISUAL INSPECTION FOR BURNT TRACE ON THE TIP SURFACE. DIELECTRIC BREAKDOWN WAS OBSERVED. THE TIP PASSED THE THERMISTOR TEST. THE FUNCTIONAL TESTING WAS NOT PERFORMED DUE TO TIP DAMAGE. THE PLANT EVALUATION IS UNDERWAY.
DURING THE EVALUATION OF THE TREATMENT TIP, SERVICE CONFIRMED DAMAGE OF ALONG THE RADIOFREQUENCY TRACE OF THE TIP. THIS EVALUATION CONFIRMS THE CUSTOMER¿S ACCOUNT OF ISSUES TO THE TIP MEMBRANE DURING THE TREATMENT. INVESTIGATION FOUND FLAME/SPARK FROM TIP MEMBRANE ARE CAUSED BY STRESS CONCENTRATIONS ON THE FLEX ASSEMBLY AT THE ADHESIVE EDGE THAT DAMAGED THE RF TRACE, CAUSING ARCING AND SUBSEQUENT BURN-THROUGH OF THE FLEX CIRCUIT MEMBRANE. DEFECTS ON THE TIP MEMBRANE CAN LEAD TO A RAISE IN TEMPERATURE OF THE TIP AND CAN POTENTIALLY CAUSE PATIENT BURNS DURING TREATMENT. THE DATACARD WAS NOT EVALUATED SO IT IS UNCLEAR IF ANY ERRORS OCCURRED THAT ALERTED OPERATOR. A REVIEW OF THE MANUFACTURING RECORDS SHOWED ALL REQUIREMENTS WERE MET. THE LOT HISTORY, TREND ANALYSIS, RISK ANALYSIS AND/OR DIRECTIONS FOR USE REVIEW WERE CONSIDERED ACCEPTABLE, WITH THE PRODUCT PERFORMING WITHIN ANTICIPATED RATES. FINAL TEST VERIFICATION SPECIFICATIONS ARE ACCEPTABLE. NO NON-CONFORMITIES OR ANOMALIES FOUND RELATED TO THIS COMPLAINT WHEN REVIEWING THE DEVICE HISTORY RECORD FOR SERIAL/LOT NUMBER (B)(6). MANUFACTURING DATE: 11/16/2020.
TREATMENT WAS ABLE TO BE COMPLETED WITH A DIFFERENT TIP.
THE INVESTIGATION IS UNDERWAY AND WILL BE UPDATED AS INFORMATION BECOMES AVAILABLE.
THE USER FACILITY STATED SHE WITNESSED A SPARK DURING TREATMENT PULSE. THERE WAS NO PATIENT CONTACT OR INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 789409 | THERMAGE CPT SYSTEM | ELECTROSURGICAL,CUTTING & COAGULATION & ACC. | GEI | SOLTA MEDICAL, INC | TTNS3.00E4-1200 | 222 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |