FDA Adverse Event Malfunction Summary report: N

RICHARD-ALLAN FINE TIP W/ RULER AND 6-UP LABELS- STERILE

MDR report key: 11893566 · Received May 27, 2021

Report

Report Number
1836161-2021-00022
Event Type
Malfunction
Date Received
May 27, 2021
Date of Event
May 5, 2021
Report Date
May 5, 2021
Manufacturer
ASPEN SURGICAL PRODUCTS, CALEDONIA
Product Code
FZZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ASPEN SURGICAL RECEIVED A REPORT FROM THE DISTRIBUTOR INDICATING THAT PRODUCT WAS FOUND WITH SEAL ISSUES. THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. PHOTOGRAPHIC EVIDENCE ALONG WITH THE MANUFACTURING LOT NUMBER WERE PROVIDED FOR REVIEW. THE DISTRIBUTOR INDICATED THAT THE DEFECTS WERE FOUND DURING INCOMING INSPECTION. A REVIEW OF THE SAMPLES CONFIRMED THE ISSUE FROM THE DISTRIBUTOR. ANALYSIS OF THE FINISHED GOOD LOT NUMBER WAS REVIEWED. NO NON-CONFORMANCE'S WERE NOTED DURING THE MANUFACTURING PROCESS. A REVIEW OF THE SAMPLE CONFIRMED THE REPORTED PROBLEM OF PART IN SEAL. THE PATIENT LABELS ARE LOADED INTO A MAGAZINE, THESE ALLOW FOR THE COMPONENT ALIGNMENT INTO THE POCKETS DURING THE DISPENSING PROCESS. DURING CONSULTATION WITH THE SUBJECT MATTER EXPERT (SME), THESE COMPONENTS CAN SLIDE OUT OF THE SEATED POCKET AND HAVE A HIGH TENDENCY FOR STATIC FRICTION LEADING TO POTENTIAL DELAY IN DISPENSING FROM THE LOADER ITSELF. THIS IS BECAUSE THE PRODUCT LIKES TO CLING TO ITSELF LEADING TO AN IMPROPER DISPENSE OF THE COMPONENT. ALTHOUGH THE ROOT CAUSE COULD NOT BE CONFIRMED, THIS THE LIKELY EVENT THAT CAUSED THE ISSUE. PRODUCTION TEAM WAS NOTIFIED OF THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

ASPEN SURGICAL RECEIVED A REPORT FROM THE DISTRIBUTOR INDICATING THAT A PEN WAS DISCOVERED WITH A SEALING ISSUE. THE ITEM WAS NOT IN USE. NO INJURY/DEATH WAS REPORTED. THIS REPORT WAS FILED UNDER OUR COMPLAINT HANDLING SYSTEM AS NUMBER (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
795233 RICHARD-ALLAN FINE TIP W/ RULER AND 6-UP LABELS- STERILE MARKING PEN FZZ ASPEN SURGICAL PRODUCTS, CALEDONIA 2630 241121

Patients

Seq Age Sex Outcome Treatment
1