AQUABEAM ROBOTIC SYSTEM

MDR report key: 11891339 · Received May 26, 2021

Report

Report Number
3012977056-2021-00035
Event Type
Injury
Date Received
May 26, 2021
Date of Event
October 21, 2020
Report Date
June 17, 2021
Manufacturer
PROCEPT BIOROBOTICS CORPORATION
Product Code
PZP
UDI-DI
B614AB20001
PMA / PMN Number
DEN170024
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

H.10 ADDITIONAL MANUFACTURER NARRATIVE: THE AQUABEAM ROBOTIC SYSTEM'S LOG FILE WAS REVIEWED, WHICH CONFIRMED NO MALFUNCTIONS DURING THE AQUABLATION PROCEDURE. THE REVIEW OF THE LOG FILE INDICATED THAT THE SYSTEM FUNCTIONED AS DESIGNED. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR AB2000-B REV. D/SERIAL NUMBER (B)(6) WAS PERFORMED, WHICH CONFIRMED THAT THERE WERE NO NONCONFORMANCES, FAILURES, DISCREPANCIES, OR MISSED STEPS DURING THE MANUFACTURING PROCESS THAT COULD BE RELATED TO THE REPORTED EVENT. THE REVIEW INDICATED THAT THE DEVICE MET ALL DESIGN AND MANUFACTURING SPECIFICATIONS WHEN RELEASED FOR DISTRIBUTION.. A REVIEW OF SIMILAR COMPLAINTS WAS CONDUCTED ON AB2000-B REV. D/SERIAL NUMBER (B)(6), WHICH CONFIRMED NO SIMILAR EVENTS. A REVIEW OF SIMILAR COMPLAINTS ACROSS ALL OTHER SYSTEMS FOUND NO OTHER SIMILAR EVENTS. THE AQUABEAM ROBOTIC SYSTEM'S INSTRUCTIONS FOR USE (IFU), IFU0101-00, REV. E, WAS REVIEWED AND STATES THE FOLLOWING: WARNINGS: PROCEDURE. AS WITH ANY SURGICAL UROLOGIC PROCEDURE, POTENTIAL PERIOPERATIVE RISKS OF THE AQUABLATION PROCEDURE INCLUDE: URETHRAL DAMAGE CAUSING FALSE PASSAGE OR STRICTURE. INFECTION. A ROOT CAUSE FOR THE REPORTED EVENT COULD NOT BE DETERMINED. THE AQUABEAM ROBOTIC SYSTEM'S IFU LISTS INFECTION AND URETHRAL DAMAGE CAUSING FALSE PASSAGE OR STRICTURE AS POTENTIAL RISKS OF THE AQUABLATION PROCEDURE. NO FURTHER INFORMATION HAS BEEN ABLE TO BE OBTAINED REGARDING THIS EVENT. BASED ON THE REVIEW OF THE LOG FILE, DHR AND IFU, THE EVENT IS CONSIDERED NOT TO BE DEVICE RELATED. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.

Additional Manufacturer Narrative · 1

ROOT CAUSE OF REPORTED EVENT HAS NOT YET BEEN ESTABLISHED. INVESTIGATION BY MANUFACTURER IS CURRENTLY INPROCESS.

Description of Event or Problem · 1

A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE ON (B)(6) 2020 FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). ON (B)(6) 2021, PROCEPT BIOROBOTICS CORPORATION BECAME AWARE THAT THE PATIENT HAD A MEATAL STENOSIS (URETHRAL STRICTURE) FOLLOWING THE AQUABLATION PROCEDURE (PER THE MANUFACTURER'S INSTRUCTIONS FOR USE, URETHRAL STRICTURE IS A POTENTIAL PERIOPERATIVE RISK OF THE AQUABLATION PROCEDURE). ON (B)(6) 2021, THE PATIENT UNDERWENT A MEATOPLASTY, AND HAS MADE A FULL RECOVERY AND IS CURRENTLY DOING GOOD. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
786986 AQUABEAM ROBOTIC SYSTEM FLUID JET REMOVAL SYSTEM PZP PROCEPT BIOROBOTICS CORPORATION AB2000 B614AB20001

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention