FDA Adverse Event Malfunction Summary report: N

GE OEC 9800

MDR report key: 1188858 · Received October 8, 2008

Report

Report Number
1720753-2008-25199
Event Type
Malfunction
Date Received
October 8, 2008
Date of Event
August 13, 2008
Report Date
August 28, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND ADJUSTED THE 5V POWER SUPPLY AND REPLACED THE BATTERIES ON THE X-RAY CONTROLLER AND SBC. SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

CUSTOMER REPORTED POOR IMAGE QUALITY. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GE OEC 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC. 9800 NA

Patients

Seq Age Sex Outcome Treatment
1