FDA Adverse Event Malfunction Summary report: N

ACT PLUS INSTRUMENT

MDR report key: 11886055 · Received May 26, 2021

Report

Report Number
2184009-2021-00045
Event Type
Malfunction
Date Received
May 26, 2021
Date of Event
May 14, 2021
Report Date
July 9, 2021
Manufacturer
PERFUSION SYSTEMS
Product Code
GKN
UDI-DI
00643169178380
PMA / PMN Number
K940426
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

CONCLUSION: THE COMPLAINT IS NOT CONFIRMED. THE ISSUE WAS NOT VERIFIED DURING SERVICE AND THE TECHNICIAN WAS UNABLE TO DUPLICATE REPORTED PROBLEM. PREVENTIVE MAINTENANCE/TESTING WAS PERFORMED PER SPECIFICATIONS. NO PATIENT/CLINICAL SAFETY ISSUES REPORTED. NO FURTHER ACTIONS TO BE TAKEN AT THIS TIME. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

ANALYSIS RESULTS - DEVICE WAS EVALUATED BY FIELD SERVICE ENGINEER AND NOT RETURNED TO MEDTRONIC FACILITY FOR ANALYSIS. THE REPORTED ERRONEOUS READINGS ISSUE WAS NOT VERIFIED DURING SERVICE. SERVICE TECHNICIAN WAS UNABLE TO DUPLICATE REPORTED PROBLEM. INSTRUMENT PASSED ACTTRAC TEST SEVERAL TIMES. PREVENTIVE MAINTENANCE WAS PERFORMED AS PER SPECIFICATION. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF THIS ACT PLUS INSTRUMENT ERRONEOUS READINGS WERE OBSERVED. USE OF THE INSTRUMENT WAS UNSPECIFIED AND THERE WAS NO REPORTED ADVERSE PATIENT EFFECT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
780280 ACT PLUS INSTRUMENT TIMER, CLOT, AUTOMATED GKN PERFUSION SYSTEMS ACT100 00643169178380

Patients

Seq Age Sex Outcome Treatment
1