FDA Adverse Event Malfunction Summary report: N

I-STAT PT/INR CARTRIDGE

MDR report key: 11885676 · Received May 26, 2021

Report

Report Number
2245578-2021-00046
Event Type
Malfunction
Date Received
May 26, 2021
Date of Event
March 22, 2021
Report Date
July 29, 2021
Manufacturer
ABBOTT POINT OF CARE
Product Code
GJS
UDI-DI
10054749000170
PMA / PMN Number
K020355
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT: # 859256. THE INVESTIGATION WAS COMPLETED ON 22-JUL-2021. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE CARTRIDGE LOT PASSED RELEASE SPECIFICATIONS. WHILE RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AG (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE), PT/INR CARTRIDGE LOT T21024 IS ASSOCIATED WITH AN UNRELATED PREVIOUSLY IDENTIFIED DEFICIENCY FOR INCREASED RATE OF QUALITY CHECK CODE 19 AND/OR STAR OUTS WHEN TESTED WITH WHOLE BLOOD, AS DOCUMENTED IN QUALITY RECORD 770729.

Additional Manufacturer Narrative · 1

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 1

ON (B)(6) 2021, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT PT/INR CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT PT/INR RESULT OF 2.6 ON A FEMALE PATIENT WITH CONGESTIVE HEART FAILURE AND A-FIB. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD: I-STAT, DATE: (B)(6) 2021, COLLECTED: 1259, TESTED:1259, PT / INR: 29.6 / 2.6; ACL TOPS, (B)(6) 2021, 1317, 1412, 48.8 / 4.2. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
779744 I-STAT PT/INR CARTRIDGE PT/INR CARTRIDGE GJS ABBOTT POINT OF CARE NA T21024 10054749000170

Patients

Seq Age Sex Outcome Treatment
1