FDA Adverse Event
Injury
Summary report: N
LIGHT CORD (SINGLE FIBER OPTIC CABLE, 10 FT).
MDR report key: 11874573
·
Received May 24, 2021
Report
- Report Number
- MW5101507
- Event Type
- Injury
- Date Received
- May 24, 2021
- Date of Event
- May 13, 2021
- Report Date
- May 20, 2021
- Manufacturer
- INVUITY, INC. / STRYKER INSTRUMENTS
- Product Code
- FST
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
INVUITY SINGLE FIBER OPTIC CABLE, 10 FT LIGHT CORD BECAME OVERHEATED IN SEVERAL AREAS DUE TO COMPROMISED FIBERS WITHIN THE CORD. THE EXTERIOR OF THE CABLE APPEARED NORMAL, HOWEVER, WHEN CONNECTED TO THE LIGHT BOX THERE WERE VISIBLE BRIGHT SPOTS THAT BECAME OVERHEATED DUE TO COMPROMISE WITHIN THE CORD. THE FIBER OPTIC CABLE WAS CONNECTED TO A RETRACTOR FOR USE DURING SURGERY. OVERHEATED AREAS OF THE CABLE RESULTED IN BURNS TO PATIENT'S CHEST. THE FIBER OPTIC CABLE WAS IMMEDIATELY REMOVED AND REPLACED WITH A NEW CORD ONCE THE OVERHEATING AND BURNS WERE NOTED. INVUITY VENDOR WAS NOTIFIED AND INDICATED THE CABLE COULD HAVE BEEN TOO TIGHTLY WOUND, THEREFORE CAUSING CORD FIBERS TO BECOME COMPROMISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 765365 | LIGHT CORD (SINGLE FIBER OPTIC CABLE, 10 FT). | LIGHT, SURGICAL, FIBEROPTIC | FST | INVUITY, INC. / STRYKER INSTRUMENTS | 19120912 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | Required Intervention |