FDA Adverse Event Injury Summary report: N

LIGHT CORD (SINGLE FIBER OPTIC CABLE, 10 FT).

MDR report key: 11874573 · Received May 24, 2021

Report

Report Number
MW5101507
Event Type
Injury
Date Received
May 24, 2021
Date of Event
May 13, 2021
Report Date
May 20, 2021
Manufacturer
INVUITY, INC. / STRYKER INSTRUMENTS
Product Code
FST
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

INVUITY SINGLE FIBER OPTIC CABLE, 10 FT LIGHT CORD BECAME OVERHEATED IN SEVERAL AREAS DUE TO COMPROMISED FIBERS WITHIN THE CORD. THE EXTERIOR OF THE CABLE APPEARED NORMAL, HOWEVER, WHEN CONNECTED TO THE LIGHT BOX THERE WERE VISIBLE BRIGHT SPOTS THAT BECAME OVERHEATED DUE TO COMPROMISE WITHIN THE CORD. THE FIBER OPTIC CABLE WAS CONNECTED TO A RETRACTOR FOR USE DURING SURGERY. OVERHEATED AREAS OF THE CABLE RESULTED IN BURNS TO PATIENT'S CHEST. THE FIBER OPTIC CABLE WAS IMMEDIATELY REMOVED AND REPLACED WITH A NEW CORD ONCE THE OVERHEATING AND BURNS WERE NOTED. INVUITY VENDOR WAS NOTIFIED AND INDICATED THE CABLE COULD HAVE BEEN TOO TIGHTLY WOUND, THEREFORE CAUSING CORD FIBERS TO BECOME COMPROMISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
765365 LIGHT CORD (SINGLE FIBER OPTIC CABLE, 10 FT). LIGHT, SURGICAL, FIBEROPTIC FST INVUITY, INC. / STRYKER INSTRUMENTS 19120912

Patients

Seq Age Sex Outcome Treatment
1 40 YR Required Intervention