FDA Adverse Event Malfunction Summary report: N

PN RELION 31G X 6MM 3B TW

MDR report key: 11872359 · Received May 24, 2021

Report

Report Number
3014704491-2021-00007
Event Type
Malfunction
Date Received
May 24, 2021
Date of Event
April 26, 2021
Report Date
June 4, 2021
Manufacturer
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
Product Code
FMI
UDI-DI
80382903206175
PMA / PMN Number
K162516
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. TESTED 7PCS RETENTION SAMPLES. NO FAILURE WAS FOUND.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PN RELION 31G X 6MM 3B TW EXPERIENCED DIFFICULT OPERATION OR WAS NOT WORKING/FUNCTIONING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 320617; BATCH NO: 0140337. CONSUMER REPORTED THAT THE PEN NEEDLES WILL NOT ATTACH TO THE PEN. CONSUMER STATED THAT HE ATTACHES THE NEEDLE BY SNAPPING THEM ON TO THE PEN, HE SAID THAT IS THE WAY THAT HE DID IT IN THE PAST. I ADVISED CONSUMER THAT THE NEEDLES MUST BE TWISTED ON TO SECURELY ATTACH RATHER THAN SNAPPING THEM ON. HE TRIED ATTACHING THE NEEDLE WHILE ON THE PHONE AND IT WORKED. DATE OF EVENT: UNKNOWN.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PN RELION 31G X 6MM 3B TW EXPERIENCED DIFFICULT OPERATION OR WAS NOT WORKING/FUNCTIONING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 320617, BATCH NO: 0140337. CONSUMER REPORTED THAT THE PEN NEEDLES WILL NOT ATTACH TO THE PEN. CONSUMER STATED THAT HE ATTACHES THE NEEDLE BY SNAPPING THEM ON TO THE PEN, HE SAID THAT IS THE WAY THAT HE DID IT IN THE PAST. I ADVISED CONSUMER THAT THE NEEDLES MUST BE TWISTED ON TO SECURELY ATTACH RATHER THAN SNAPPING THEM ON. HE TRIED ATTACHING THE NEEDLE WHILE ON THE PHONE AND IT WORKED. DATE OF EVENT: UNKNOWN.