ACHIEVE MAPPING CATHETER - 20 MM
Report
- Report Number
- 9617601-2021-00104
- Event Type
- Death
- Date Received
- May 24, 2021
- Date of Event
- May 17, 2021
- Report Date
- June 14, 2021
- Manufacturer
- MEDTRONIC MEXICO S. DE R.L. DE CV
- Product Code
- DRF
- PMA / PMN Number
- K102588
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE DATA FILES WERE RETURNED AND ANALYZED. THE DATA FILES SHOWED TWO APPLICATIONS WERE PERFORMED WITH AN AFAPRO28 BALLOON CATHETER WITH LOT: 14021 ON THE DAY OF THE EVENT. A PERSISTENT SYSTEM NOTICE#: 50005 (INDICATING THAT THE SAFETY SYSTEM DETECTED FLUID IN THE CATHETER AND STOPPED THE INJECTION) WAS RECEIVED. IN CONCLUSION, THE CLINICAL ISSUES (DEATH, EFFUSION, PERFORATION) OCCURRED DURING THE PROCEDURE. THERE IS NO INDICATION OF A RELATION OF THE ADVERSE EVENT TO THE PERFORMANCE AND MALFUNCTION OF THE PRODUCTS. THE PHYSICAL PRODUCTS WERE DISCARDED AND ARE NOT AVAILABLE FOR ANALYSIS. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, THE PATIENT BEGAN SHOWING SIGNS OF DISTRESS AND AN EMERGENCY THORACOTOMY WAS PERFORMED. A PERFORATION WAS NOTED AT THE ROOF OF THE LEFT ATRIUM AND A PERICARDIAL EFFUSION OBSERVED. THE CASE WAS ABORTED. THE PATIENT DID NOT SURVIVE THE INTERVENTIONS. THE PATIENT IS DECEASED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 765575 | ACHIEVE MAPPING CATHETER - 20 MM | CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT | DRF | MEDTRONIC MEXICO S. DE R.L. DE CV | 990063-020 | 220733863 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Death | AFAPRO28 BALLOON CATHETER, 4FC12 SHEATH.| AFAPRO28 BALLOON CATHETER, 4FC12 SHEATH |