LUMIRADX SARS-COV-2 AG TEST STRIP KIT
Report
- Report Number
- 3012642695-2021-01137
- Event Type
- Malfunction
- Date Received
- May 24, 2021
- Date of Event
- May 1, 2021
- Report Date
- May 24, 2021
- Manufacturer
- LUMIRADX UK LTD.
- Product Code
- QKP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 117
Narratives
PATIENT WAS REPORTED TO BE SYMPTOMATIC WHEN INITIAL LUMIRADX SARS-COV-2 AG TEST WAS PERFORMED ON (B)(6) 2021 WITH STRIP LOT 6000164 AND INSTRUMENT SERIAL NUMBER (B)(4). SECONDARY TESTING WAS PERFORMED BY THE PATIENT WITH TWO TAKE-HOME TESTS FROM ANOTHER MANUFACTURER'S MAKE. BOTH RESULTS WERE NEGATIVE. CONFIRMATORY TESTING WAS PERFORMED TWICE VIA POLYMERASE CHAIN REACTION (PCR). BOTH CONFIRMATORY TEST RESULTS WERE NEGATIVE. THE CUSTOMER REPORTED THE FOLLOWING PROCESS FOR TESTING ON THE LUMIRADX PLATFORM: "THIS PATIENT'S SAMPLE WAS COLLECTED BY THE NURSE." THE CUSTOMER REPORTED THEIR CLEANING PRACTICES TO BE "PDI BLEACH WIPES- CLEAN AFTER EACH POSITIVE TEST AND AT THE END OF THE DAY.", AND FURTHER REPORTED THAT GLOVES ARE CHANGED AFTER EACH SAMPLE. LOT 6000164 MET ALL DEFINED QC CRITERIA AT THE TIME IT WAS RELEASED AND TESTING USING NEGATIVE NASAL SWABS FROM IN-HOUSE DONORS MEETS EXPECTED PERFORMANCE FOR USE IN THE FIELD. REVIEW OF PRODUCT RISK ASSESSMENT SARS-COV-2 AG ASSAY REVISION 7, RESULTED IN SEVERITY OF MODERATE, AS FOLLOWS: DELAYED DIAGNOSIS LEADING TO DELAYED TREATMENT OF THE TRUE CAUSE OF THE PATIENT'S SYMPTOMS. UNNECESSARY ISOLATION. NO PATIENT HARM, INJURY OR ADVERSE HEALTH CONSEQUENCES WERE COMMUNICATED BY THE CUSTOMER TO LUMIRADX FOR THE REPORTED DISCORDANT RESULT. TRENDING DATA FOR DISCORDANT RESULTS WAS REVIEWED FOR THIS LOT AND THE OCCURRENCE RATE PER QUANTITY OF STRIPS IN THE FIELD WAS CALCULATED AS (B)(4). LUMIRADX SARS-COV-2 AG TEST PRODUCT INSERT CLAIMS A SPECIFICITY OF 96.6% WITH A REFERENCE RT-PCR ASSAY AND IT IS ACCEPTED THAT UP TO 3.4% OF TEST STRIPS MAY GENERATE A DISCORDANT FALSE POSITIVE RESULT. ROOT CAUSE DETERMINATION: NO DEFINITIVE ROOT CAUSE HAS BEEN DETERMINED FOR THE REPORTED DISCORDANT RESULT BASED ON THE INFORMATION CURRENTLY AVAILABLE. INVESTIGATION CONCLUSION AND STATUS OF INVESTIGATION: NO FURTHER INVESTIGATION IS CONSIDERED NECESSARY AT THIS TIME. THIS STRIP LOT CONTINUES TO DEMONSTRATE FIELD PERFORMANCE WITHIN SPECIFICATION OF PRODUCT CLAIMS RELATIVE TO THE QUANTITY OF STRIPS FROM THIS LOT IN THE FIELD. REPORTS OF DISCORDANT RESULTS FOR THIS LOT WILL CONTINUE TO BE TRENDED AND REVIEWED, WITH ACTION TAKEN AS APPROPRIATE IN RESPONSE TO ANY ADVERSE TRENDS OR EVENTS INCLUDING A FOLLOW-UP REPORT.
THE SYMPTOMATIC PATIENT REPORTED A SUSPECTED FALSE (INCORRECT) POSITIVE RESULT FROM A RAPID RESULT COVID TEST SYSTEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 765567 | LUMIRADX SARS-COV-2 AG TEST STRIP KIT | CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM | QKP | LUMIRADX UK LTD. | 6000164 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |