LCS COMP RP APG INS TRL SM+ 10
Report
- Report Number
- 1818910-2021-10994
- Event Type
- Malfunction
- Date Received
- May 24, 2021
- Date of Event
- May 11, 2021
- Report Date
- May 11, 2021
- Manufacturer
- DEPUY ORTHOPAEDICS INC US
- Product Code
- HWT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT # : (B)(4). INVESTIGATION SUMMARY : EXAMINATION OF THE RETURNED DEVICE CONFIRMED THE REPORTED DAMAGE. THE NOTED DAMAGE IS CONSISTENT WITH MATERIAL OVERLOAD THROUGH THE USE OF EXCESSIVE FORCE AND THE INVESTIGATION DID NOT ESTABLISH A NEED FOR CORRECTIVE ACTION. DEPUY SYNTHES CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION MAY BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT : THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). WHERE THE PRODUCT AND LOT CODE WAS PROVIDED, A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED.
PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
ADDITIONAL INFORMATION RECEIVED STATES THAT ALL PIECES OF THE BROKEN INSTRUMENT WERE RETRIEVED FROM THE PATIENT.
PRODUCT COMPLAINT #: (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT ON (B)(6) 2021, DURING TKA SURGERY, WHEN THE SURGEON TRIED TO REMOVE THE TRIAL INSERT CAREFULLY AFTER INSERTING IT, THE TRIAL INSERT BROKE FROM THE MIDDLE. THE TRIAL INSERT WAS NOT USED FOR SURGERY BECAUSE OF EVENTS DURING THE TRIAL. ALL PIECES WERE RETRIEVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 769658 | LCS COMP RP APG INS TRL SM+ 10 | KNEE INSTRUMENT : TIBIAL TRIALS | HWT | DEPUY ORTHOPAEDICS INC US |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |