FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 11.5MM

MDR report key: 11868092 · Received May 22, 2021

Report

Report Number
0002023141-2021-01370
Event Type
Injury
Date Received
May 22, 2021
Date of Event
March 4, 2021
Report Date
September 22, 2021
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019577
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4) DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1232150). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1232150) FOR SIMILAR CATEGORY MEDICAL: INFECTION THROUGH THE MANUFACTURERS INTERNAL SYSTEM AND NO OTHER COMPLAINT WAS IDENTIFIED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET (B)(4). PATIENT WEIGHT: NOT PROVIDED. FAX NUMBER: NOT PROVIDED. PMA/510(K) NUMBER: ADDITIONAL NUMBER: K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR INFECTION EVENTS IDENTIFYING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO THE WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED) FACTORS POTENTIALLY IMPACTING THE STERILITY OF THE IMPLANT AND ITS ENVIRONMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE INFECTION EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED IMPLANT AT TOOTH LOCATION # 29 WAS REMOVED DUE TO INFECTION, ABSCESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
764547 IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 11.5MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVH11 1232150 00889024019577

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention