FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 0.5ML 31GA 6MM

MDR report key: 11867639 · Received May 21, 2021

Report

Report Number
1920898-2021-00585
Event Type
Malfunction
Date Received
May 21, 2021
Date of Event
April 24, 2021
Report Date
June 18, 2021
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: CUSTOMER RETURNED TWO IMAGES OF A SINGLE SYRINGE WITH A POLYBAG IDENTIFYING IT AS A 0.5ML, 31 GAUGE, 6MM SYRINGE FROM LOT 0083432. THE SYRINGE HAS HAD ITS NEEDLE SHIELD AND HUB SEPARATE FROM THE BARREL. THE HUB HAS BECOME LODGED INSIDE THE SHIELD. THERE IS NO DAMAGE TO EITHER THE CONNECTOR AT THE DISTAL TIP OF THE BARREL OR ITS RESPECTIVE NEEDLE HUB. NO SIGNS OF USE AND NO OTHER DEFECTS FOUND. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 0083432. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE FOUR (4) NOTIFICATIONS NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. BASED ON THE IMAGES RECEIVED, BD WAS ABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE OF NEEDLE HUB SEPARATION. CAPA (B)(4) HAS BEEN OPENED TO ADDRESS THIS ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USE WITH BD SYRINGE 0.5ML 31GA 6MM THE HUB WAS SEPARATED FROM DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE TO ENGLISH: "UNFORTUNATELY AGAIN, I COME TO INFORM YOU ABOUT A SYRINGE THAT CAME WITHOUT A NEEDLE".

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO USE WITH BD SYRINGE 0.5ML 31GA 6MM THE HUB WAS SEPARATED FROM DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "UNFORTUNATELY AGAIN, I COME TO INFORM YOU ABOUT A SYRINGE THAT CAME WITHOUT A NEEDLE".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
763237 BD SYRINGE 0.5ML 31GA 6MM PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 0083432

Patients

Seq Age Sex Outcome Treatment
1