BD INTIMA-II CLOSED IV CATHETER SYSTEM
Report
- Report Number
- 3006948883-2021-00586
- Event Type
- Malfunction
- Date Received
- May 21, 2021
- Date of Event
- January 16, 2021
- Report Date
- May 24, 2021
- Manufacturer
- BD (SUZHOU)
- Product Code
- FOZ
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9233892. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT THE BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM LEAKED BLOOD FROM THE NEEDLE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: "WHEN THE PUNCTURE WAS SUCCESSFUL AND THE STEEL NEEDLE WAS WITHDRAWN, BLOOD LEAKAGE APPEARED AT THE TAIL OF THE INDWELLING NEEDLE AND A NEW INDWELLING NEEDLE WAS REPLACED".
INITIAL REPORTER PHONE#: (B)(6). A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM LEAKED BLOOD FROM THE NEEDLE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: "WHEN THE PUNCTURE WAS SUCCESSFUL AND THE STEEL NEEDLE WAS WITHDRAWN, BLOOD LEAKAGE APPEARED AT THE TAIL OF THE INDWELLING NEEDLE AND A NEW INDWELLING NEEDLE WAS REPLACED."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 762591 | BD INTIMA-II CLOSED IV CATHETER SYSTEM | INTRAVASCULAR CATHETER | FOZ | BD (SUZHOU) | 9233892 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |