HEARTMATE MOBILE POWER UNIT, UK
Report
- Report Number
- 2916596-2021-02557
- Event Type
- Malfunction
- Date Received
- May 21, 2021
- Date of Event
- April 29, 2021
- Report Date
- June 30, 2021
- Manufacturer
- THORATEC CORPORATION
- Product Code
- DSQ
- PMA / PMN Number
- P160054
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- 501
Narratives
MANUFACTURER'S INVESTIGATION CONCLUSION: THE REPORTED EVENT OF A NO EXTERNAL POWER ALARM WAS NOT CONFIRMED. THE SUBMITTED LOG FILE BELONGED TO HSC-033791 WHICH WAS THE CONTROLLER THE PATIENT EXCHANGED TO. THERE WAS NO LOG FILE FOR THE CONTROLLER IN QUESTION. THE MPU, SERIAL (B)(6), WAS NOT RETURNED FOR ANALYSIS AND REMAINS IN USE. THE PROVIDED INFORMATION STATED THAT EXCHANGING THEIR CONTROLLER RESOLVED THE NO EXTERNAL POWER ALARM. ADDITIONAL INFORMATION PROVIDED ON 14JUN21 STATED THAT THE POWER CORD CAME LOOSE AT THE WALL. A ROOT CAUSE FOR THE REPORTED EVENT WAS NOT CONCLUSIVELY DETERMINED THROUGH THIS ANALYSIS. DEVICE HISTORY RECORDS WAS REVIEWED AND SHOWED NO DEVIATIONS FROM MANUFACTURING OR QUALITY ASSURANCE SPECIFICATIONS. THE PATIENT HANDBOOK CAUTIONS THE USERS TO CALL THEIR HOSPITAL CONTACTS IF THEY THINK, FOR ANY REASON, ANY PORTION OF THEIR EQUIPMENT IS NOT FUNCTIONING AS USUAL, IS BROKEN, OR THEY ARE UNCOMFORTABLE WITH THE OPERATION OF THE EQUIPMENT. HEARTMATE 3 INSTRUCTIONS FOR USE (IFU) SECTION 7-¿ALARMS AND TROUBLESHOOTING¿ AND HEARTMATE 3 PATIENT HANDBOOK SECTION 5-¿ALARMS AND TROUBLESHOOTING¿ EXPLAIN HOW TO PROPERLY INTERPRET AND TROUBLESHOOT NO EXTERNAL POWER ALARMS. THE "PATIENT CARE MANAGEMENT" SECTION OF THE IFU INSTRUCTS THE PATIENT TO KEEP A FLASHLIGHT, FULLY-CHARGED BATTERIES, AND BATTERY CLIPS WITHIN REACH TO BE PREPARED FOR A POWER OUTAGE. NO FURTHER INFORMATION WAS PROVIDED. THE MANUFACTURER IS CLOSING THE FILE ON THIS EVENT.
RELATED MANUFACTURER REPORT NUMBERS: 2916596-2021-02555 AND 2916596-2021-02556.
THE RESULTS/METHOD AND CONCLUSION CODES ALONG WITH INVESTIGATION RESULTS WILL BE PROVIDED IN THE FINAL REPORT.
IT WAS REPORTED THAT WHEN THE PATIENT SWITCHED TO THE MPU FROM BATTERY, THE RUNNING SYMBOL WAS OFF. THERE WAS A NO EXTERNAL POWER AND LOW FLOW ALARM OCCURRED AT THE SAME TIME. THE PATIENT WAS CLINICALLY STABLE DURING THE EVENT. THE PATIENT EXCHANGED THEIR CONTROLLER AND THE ALARMS RESOLVED, EXCEPT THE LOW FLOW. THE PATIENT WAS ASYMPTOMATIC DURING THE EXCHANGE. AN ECHO WAS PERFORMED WHICH FOUND NO PUMP MALPOSITION OR OBSTRUCTION. THE LOW FLOWS HAVE CONTINUED INTERMITTENTLY. THE PLAN OF CARE WAS TO CONTINUE WITH ROUTINE CARE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 761705 | HEARTMATE MOBILE POWER UNIT, UK | VENTRICULAR (ASSISST) BYPASS | DSQ | THORATEC CORPORATION | 107758UK | 7054860 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Required Intervention |