FDA Adverse Event Malfunction Summary report: N

CS300

MDR report key: 11864033 · Received May 21, 2021

Report

Report Number
2249723-2021-01075
Event Type
Malfunction
Date Received
May 21, 2021
Date of Event
April 29, 2021
Report Date
August 2, 2021
Manufacturer
DATASCOPE CORP. - MAHWAH
Product Code
DSP
UDI-DI
10607567107882
PMA / PMN Number
K063525
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

HISTORICAL DATA ANALYSIS: (4109/131) THE REVIEW OF THE HISTORICAL DATA INDICATES THAT NO OTHER SIMILAR COMPLAINT WAS REPORTED FOR THE SAME SERIAL NUMBER AND REPORTED FAILURE MODE. TREND ANALYSIS: (4110/131) THE OVERALL 24 MONTH PRODUCT COMPLAINT TREND DATA FOR THE PERIOD ((B)(6) 2019 THROUGH (B)(6) 2021) WAS REVIEWED. THERE WERE NO TRIGGERS IDENTIFIED FOR THE REVIEW PERIOD. ANALYSIS OF PRODUCTION: (3331/131) THE DEVICE HISTORY RECORDS REVIEW CONCLUDED THAT THERE WERE NO NCMRS, REWORK, OR DEVIATIONS DOCUMENTED FOR THE REPORTED SERIAL NUMBER. BASED ON THE DHR/LHR REVIEW RESULTS, IT WAS DETERMINED THAT THERE IS NO RELATION BETWEEN THE MANUFACTURING PROCESS AND THE REPORTED FAILURE.

Description of Event or Problem · 0

N/A.

Additional Manufacturer Narrative · 1

THE GETINGE SERVICE TERRITORY MANAGER (STM) THAT ENCOUNTERED THE ISSUE, REPORTED THAT BATTERIES ONLY PROVIDED POWER TO THE DEVICE FOR 92 MINUTES. THE STM REPLACED BATTERIES AND COMPLETE PM WITH FULL CALIBRATION. UNIT PASSED ALL FUNCTIONAL AND SAFETY TESTS PER FACTORY SPECIFICATIONS. THE IABP WAS RETURNED TO THE CUSTOMER AND CLEARED FOR CLINICAL USE. (B)(6).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE (PM) PERFORMED BY A GETINGE SERVICE TERRITORY MANAGER (STM), THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) FAILED BATTERY SHORT RUN TIME TEST. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
760443 CS300 SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. - MAHWAH 0998-00-3023-53 N/A 10607567107882

Patients

Seq Age Sex Outcome Treatment
1 N/A.