Narratives
Description of Event or Problem · 1
PATIENT BECAME BRADYCARDIC/HYPERTENSIVE SUDDENLY AROUND 1430, IT WAS FOUND THAT THE WATER HEATER ON THE ECMO CIRCUIT HAD BEEN PUT ONTO STANDBY UNKNOWINGLY, UPON DISCOVERING THIS IT WAS RESTARTED, PATIENT WAS REWARMED SLOWLY AND BRADYCARDIA/ HYPERTENSION RESOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 760622 | CARDIOQUIP MODULAR COOLER-HEATER (MCH) | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | CARDIOQUIP, LLC | MCH-1000(I) |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 30 DA |