VITEK MS INSTRUMENT
Report
- Report Number
- 9615754-2021-00165
- Event Type
- Malfunction
- Date Received
- May 21, 2021
- Report Date
- May 21, 2021
- Manufacturer
- BIOMERIEUX, SA
- Product Code
- QBN
- UDI-DI
- 03573026359119
- PMA / PMN Number
- K181412
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
CONTEXT: A CUSTOMER IN (B)(6) NOTIFIED BIOMÉRIEUX OF OBTAINING A MISIDENTIFICATION OF WICKERHAMOMYCES ANOMALUS AS CANDIDA PELLICULOSA WHEN USING THE VITEK® MS SYSTEM, REFERENCE 410895, SERIAL NUMBER #(B)(4). INVESTIGATION RESULTS: FINE TUNING: FINE TUNING WAS NOT WITHIN SPECIFICATION, SEVERAL MANDATORY CRITERIA WERE NOT ACHIEVED. SPOT PREPARATION QUALITY: THE CALIBRATOR ¿ALL PEAKS¿ VALUES ARE QUITE HETEROGENEOUS AND, THEREFORE, NOT OPTIMAL. THE SAMPLE ¿ALL PEAKS¿ ARE HOMOGENEOUS. KNOWLEDGE BASE REVIEW: WICKERHAMOMYCES ANOMALUS IS THE SEXUAL FORM (TELEOMORPH) OF CANDIDA PELLICULOSA. THE EXPECTED IDENTIFICATION, WICKERHAMOMYCES ANOMALUS, IS PRESENT UNDER THE NAME OF CANDIDA PELLICULOSA (ASEXUAL FORM ¿ ANAMORPH). SAMPLE DATA ANALYSIS: ANALYSIS OF SPECIMEN DATA FILES DETERMINED THAT THE NUMBER OF ALL PEAKS ARE HOMOGENEOUS DURING THE ISSUE. THE MISIDENTIFICATION WAS OBTAINED WITH AN ACCEPTABLE IDENTIFICATION SCORE. LOCAL CUSTOMER SERVICE RECOMMENDED ACTIONS : - CHECK SAMPLE PREPARATION WITH THE CUSTOMER TO HELP HIM TO IMPROVE HIS SPOT PREPARATION TECHNIQUE (VIDEO, TRAINING, ¿ ) - PERFORM A NEW FINE TUNING RESPECTING THE FINE TUNING PROCEDURE. CONCLUSION : A SYSTEM MALFUNCTION HAS NOT BEEN CONFIRMED. VITEK MS GAVE THE EXPECTED IDENTIFICATION AS CANDIDA PELLICULOSA IS THE ASEXUAL FORM (ANAMORPH) OF WICKERHAMOMYCES ANOMALUS. NO PROBLEM HAS BEEN DETECTED REGARDING THE STRAIN : WICKERHAMOMYCES ANOMALUS.
INTENDED USE. THE VITEK® MS SYSTEM, REFERENCE 410895, IS A MASS SPECTROMETER USING MALDI-TOF (MATRIX-ASSISTED LASER DESORPTION/IONIZATION-TIME OF FLIGHT) TECHNOLOGY FOR THE IDENTIFICATION OF MICROORGANISMS CULTURED FROM CLINICAL SPECIMENS. DESCRIPTION OF THE ISSUE: A CUSTOMER IN (B)(6) NOTIFIED BIOMÉRIEUX OF OBTAINING A MISIDENTIFICATION OF WICKERHAMOMYCES ANOMALUS AS CANDIDA PELLICULOSA WHEN USING THE VITEK® MS SYSTEM, REFERENCE 410895, SERIAL NUMBER #(B)(4). IDENTIFICATION OF THE STRAIN USING DNA SEQUENCING METHOD OBTAINED WICKERHAMOMYCES ANOMALUS. WICKERHAMOMYCES ANOMALUS IS NOT IN THE VITEK MS KB 3.0 DATABASE; VITEK MS PROVIDED AN INCORRECT ORGANISM ID TO CANDIDA PELLICULOSA. NO INFORMATION HAS BEEN PROVIDED BY THE CUSTOMER REGARDING ANY POTENTIAL PATIENT IMPACT OF THE MISIDENTIFICATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 759100 | VITEK MS INSTRUMENT | VITEK MS INSTRUMENT | QBN | BIOMERIEUX, SA | 410895 | 03573026359119 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |