FDA Adverse Event Malfunction Summary report: N

FS DISPOSABLE INTERFACE

MDR report key: 11857726 · Received May 20, 2021

Report

Report Number
3006695864-2021-07773
Event Type
Malfunction
Date Received
May 20, 2021
Date of Event
March 4, 2021
Report Date
July 6, 2021
Manufacturer
AMO MANUFACTURING USA, LLC
Product Code
GEX
UDI-DI
15050474534688
PMA / PMN Number
K060372
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

SECTION D9: DEVICE AVAILABLE FOR EVALUATION: YES, RETURNED TO MANUFACTURER ON 05/04/2021. SECTION H3: DEVICE EVALUATED BY MANUFACTURER: YES. DEVICE EVALUATION: A VISUAL INSPECTION OF THE RETURNED PRODUCT DID NOT REVEAL ANY OBVIOUS DEFECTS OR DAMAGE. FUNCTIONAL TESTING WAS PERFORMED AND REPORTED ISSUE NOT CONFIRMED. A RECORD REVIEW WAS PERFORMED. A PRODUCT DEFICIENCY REVIEW WAS PERFORMED AND THERE IS NO PRODUCT DEFICIENCY IDENTIFIED. A DOCUMENT, LOT HISTORY, AND TRENDING WERE REVIEWED. THERE IS NOT A RECOGNIZABLE ADVERSE TREND. THE RISKS AND MITIGATIONS ASSOCIATED WITH THE COMPLAINT ISSUE ARE IDENTIFIED IN EXISTING RISK DOCUMENTS AND NO NEW RISKS WERE IDENTIFIED AS PART OF THIS INVESTIGATION. THE REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR THE PATIENT INTERFACE WAS PERFORMED. THE DEVICE AND ITS COMPONENTS MET ALL SPECIFICATIONS PRIOR TO BEING RELEASED. MANUFACTURING HAS BEEN RULED OUT AS A POTENTIAL CAUSE FOR THE REPORTED ISSUE. BASED ON THE INVESTIGATION RESULTS, NO CORRECTIVE ACTION HAS BEEN ISSUED. CONCLUSION: BASED ON THE INFORMATION OBTAINED, PRODUCT MALFUNCTION AND PRODUCT DEFICIENCY CANNOT BE CONFIRMED. NO FURTHER INVESTIGATION IS REQUIRED ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 1

THE PATIENT INTERFACE (PI) SUCTION RING MAY LOSE SUCTION DURING A PROCEDURE. LABEL COPY STATES CORNEAL FIXATION VACUUM LOSS CAN OCCUR. THERE ARE SEVERAL FACTORS THAT MAY CONTRIBUTE TO SUCTION ISSUES SUCH AS DOCTOR¿S TECHNIQUE IN APPLYING THE SUCTION RING TO THE CORNEA, DOCTOR¿S TECHNIQUE IN SQUEEZING THE PI CLIP TO SECURE THE SUCTION RING TO THE PI CONE AND PATIENT ANATOMY AFFECTING THE INTERFACE BETWEEN THE PATIENT¿S CORNEA AND THE SUCTION RING. A REVIEW OF RECORDS RELATED TO THE DEVICE INCLUDING LABELING, COMPLAINT TRENDING, AND RISK DOCUMENTATION WILL BE PERFORMED. UPON COMPLETION OF THIS REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION OBTAINED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SUCTION LOSS DURING LASER FIRE OCCURRED USING THE PATIENT INTERFACE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT INJURY AND/OR COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
752243 FS DISPOSABLE INTERFACE POWERED LASER SURGICAL INSTRUMENT GEX AMO MANUFACTURING USA, LLC 590106AN 60231273 15050474534688

Patients

Seq Age Sex Outcome Treatment
1 FEMTO SN (B)(6).