HAKIM PERITON CATH,120CM
Report
- Report Number
- 3013886523-2021-00227
- Event Type
- Injury
- Date Received
- May 20, 2021
- Report Date
- June 11, 2021
- Manufacturer
- INTEGRA LIFESCIENCES SWITZERLAND SAR
- Product Code
- JXG
- PMA / PMN Number
- K944222
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THE CATHETER WAS NOT RETURNED FOR EVALUATION (AS PER CUSTOMER, PRODUCT NOT AVAILABLE) AND NO LOT NUMBER INFORMATION HAS BEEN PROVIDED; THEREFORE, AN EVALUATION OF THE DEVICE COULD NOT BE PERFORMED, AND MANUFACTURING RECORDS COULD NOT BE REVIEWED. THE CAUSE(S) OF THE DIFFICULTY REPORTED BY THE CUSTOMER COULD NOT BE DETERMINED. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE IN THE FUTURE, THIS COMPLAINT WILL BE REOPENED, AND THE RESPECTIVE EVALUATION PERFORMED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR ISSUES. AT PRESENT, WE CONSIDER THIS COMPLAINT TO BE CLOSED.
N/A.
AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.
2 OF 2 REPORTS. OTHER MFG REPORT NUMBER: 3013886523-2021-00226. A PHYSICIAN REPORTED THAT IN 2001, A CODMAN HAKIM PROGRAMMABLE VALVE WAS IMPLANTED TO A (B)(6) YEAR-OLD FEMALE PATIENT VIA V-P SHUNT WITH A SETTING OF 100 MMH20. ON (B)(6) 2021, THE PATIENT COMPLAINED OF A HEADACHE. CONFIRMATION VIA MRI SHOWED THAT THE CATHETER HAD FALLEN OFF AND A SLIT-LIKE VENTRICLE WAS NOTED. ONLY THE VALVE WAS REPLACED ON (B)(6) 2021. THE IMPLANTED ABDOMINAL CATHETER WAS EXTENDED BY USING A CONNECTOR (823053) AND BACTISEAL (823073). THE CAUSE OF THE DEVIATION OF THE ABDOMINAL CATHETER WAS THE INCREASE IN HEIGHT OF THE PATIENT AND THE OCCURRENCE OF GRAVITATIONAL FORCE. THE PATIENT¿S CONDITION HAD IMPROVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 755979 | HAKIM PERITON CATH,120CM | STANDARD CATHETERS | JXG | INTEGRA LIFESCIENCES SWITZERLAND SAR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR |