TRACK MOUNTED DENTAL LIGHT
Report
- Report Number
- 1017522-2021-00001
- Event Type
- Injury
- Date Received
- May 20, 2021
- Date of Event
- October 21, 2020
- Report Date
- June 23, 2021
- Manufacturer
- KAVO DENTAL TECHNOLOGIES, LLC
- Product Code
- EAZ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- DENTIST
Narratives
THE DENTIST RETURNED THE DENTAL LIGHT TO KAVO DENTAL TECHNOLOGIES, LLC (DBA PELTON & CRANE) ON (B)(6) 2021 FOR EVALUATION. AS A RESULT OF THE EVALUATION IT WAS DETERMINED THAT THE ROLL PIN AND 2 OF 4 SET SCREWS WERE NOT PROPERLY INSTALLED BY THE DISTRIBUTOR DURING INSTALLATION. THE ROLL PINS AND SET SCREWS WILL PREVENT THE LIGHT FROM UNSCREWING FROM THE TROLLEY AND/OR POST AFTER INSTALLATION. THE PELTON & CRANE INSTALLATION INSTRUCTIONS CLEARLY STATES TO PROPERLY INSTALL THE ROLL PINS AND SET SCREWS DURING INSTALLATION OF THE TRACK LIGHT. THE INSTALLATION INSTRUCTIONS ALSO LISTS WARNINGS TO ENSURE THE ROLL PINS AND SET SCREWS ARE INSTALLED. (B)(6) (PELTON & CRANE QUALITY SPECIALIST) CONTACTED THE DR'S OFFICE ON (B)(6) 2021 AND SPOKE WITH DANIELLE WITH THE FRONT OFFICE, AND INFORMED DANIELLE THAT THE CAUSE OF FAILURE WAS DUE TO IMPROPER INSTALLATION BY HER DEALER. THIS CONCLUDES THE INVESTIGATION.
THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION TO DATE. ATTEMPTS ARE BEING MADE TO HAVE THE LIGHT RETURNED FOR EVALUATION AND IF THE DEVICE IS RETURNED, AN EVALUATION WILL BE PERFORMED, AND A FOLLOW UP REPORT SUBMITTED.
IT WAS REPORTED THAT AT THE TIME OF THE EVENT, A PATIENT WAS IN THE ROOM IN THE DENTAL CHAIR AND THE DENTAL ASSISTANT WAS STANDING BEHIND THE PATIENT. THE LIGHT WAS SEMI FOLDED AND TO THE SIDE OF THE PATIENT, THE LIGHT FELL STRIKING THE DENTAL ASSISTANT ON THE RIGHT SHOULDER AND RIGHT HAND. THE LIGHT DID NOT HIT THE PATIENT AND THE LIGHT WAS NOT IN USE AT THE TIME OF THE INCIDENT. THE LIGHT BROKE/CAME APART AT THE ELBOW/BEND IN THE MECHANICAL ARM. ONCE THE LIGHT FELL AND HIT THE FLOOR, THE HEAD BROKE AT THE HANDLE/YOKE. THE DENTAL ASSISTANT WENT TO THE LOCAL URGENT CARE FOR TREATMENT AND HER ARM WAS PUT IN A SPLINT. THE DENTAL ASSISTANT CHOSE NOT TO RETURN TO WORK AFTER THE INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 752368 | TRACK MOUNTED DENTAL LIGHT | DENTAL LIGHT | EAZ | KAVO DENTAL TECHNOLOGIES, LLC | HL3T |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |