TRACK MOUNTED DENTAL LIGHT

MDR report key: 11854243 · Received May 20, 2021

Report

Report Number
1017522-2021-00001
Event Type
Injury
Date Received
May 20, 2021
Date of Event
October 21, 2020
Report Date
June 23, 2021
Manufacturer
KAVO DENTAL TECHNOLOGIES, LLC
Product Code
EAZ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 0

THE DENTIST RETURNED THE DENTAL LIGHT TO KAVO DENTAL TECHNOLOGIES, LLC (DBA PELTON & CRANE) ON (B)(6) 2021 FOR EVALUATION. AS A RESULT OF THE EVALUATION IT WAS DETERMINED THAT THE ROLL PIN AND 2 OF 4 SET SCREWS WERE NOT PROPERLY INSTALLED BY THE DISTRIBUTOR DURING INSTALLATION. THE ROLL PINS AND SET SCREWS WILL PREVENT THE LIGHT FROM UNSCREWING FROM THE TROLLEY AND/OR POST AFTER INSTALLATION. THE PELTON & CRANE INSTALLATION INSTRUCTIONS CLEARLY STATES TO PROPERLY INSTALL THE ROLL PINS AND SET SCREWS DURING INSTALLATION OF THE TRACK LIGHT. THE INSTALLATION INSTRUCTIONS ALSO LISTS WARNINGS TO ENSURE THE ROLL PINS AND SET SCREWS ARE INSTALLED. (B)(6) (PELTON & CRANE QUALITY SPECIALIST) CONTACTED THE DR'S OFFICE ON (B)(6) 2021 AND SPOKE WITH DANIELLE WITH THE FRONT OFFICE, AND INFORMED DANIELLE THAT THE CAUSE OF FAILURE WAS DUE TO IMPROPER INSTALLATION BY HER DEALER. THIS CONCLUDES THE INVESTIGATION.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION TO DATE. ATTEMPTS ARE BEING MADE TO HAVE THE LIGHT RETURNED FOR EVALUATION AND IF THE DEVICE IS RETURNED, AN EVALUATION WILL BE PERFORMED, AND A FOLLOW UP REPORT SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AT THE TIME OF THE EVENT, A PATIENT WAS IN THE ROOM IN THE DENTAL CHAIR AND THE DENTAL ASSISTANT WAS STANDING BEHIND THE PATIENT. THE LIGHT WAS SEMI FOLDED AND TO THE SIDE OF THE PATIENT, THE LIGHT FELL STRIKING THE DENTAL ASSISTANT ON THE RIGHT SHOULDER AND RIGHT HAND. THE LIGHT DID NOT HIT THE PATIENT AND THE LIGHT WAS NOT IN USE AT THE TIME OF THE INCIDENT. THE LIGHT BROKE/CAME APART AT THE ELBOW/BEND IN THE MECHANICAL ARM. ONCE THE LIGHT FELL AND HIT THE FLOOR, THE HEAD BROKE AT THE HANDLE/YOKE. THE DENTAL ASSISTANT WENT TO THE LOCAL URGENT CARE FOR TREATMENT AND HER ARM WAS PUT IN A SPLINT. THE DENTAL ASSISTANT CHOSE NOT TO RETURN TO WORK AFTER THE INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
752368 TRACK MOUNTED DENTAL LIGHT DENTAL LIGHT EAZ KAVO DENTAL TECHNOLOGIES, LLC HL3T

Patients

Seq Age Sex Outcome Treatment
1 Other