FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 11852924 · Received May 19, 2021

Report

Report Number
1645337-2021-05612
Event Type
Injury
Date Received
May 19, 2021
Date of Event
February 16, 2021
Report Date
May 3, 2021
Manufacturer
MENTOR MEDICAL SYSTEM B.V.
Product Code
FTR
UDI-DI
00081317000341
PMA / PMN Number
P030053
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ON (B)(6) 2021, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ADDITIONALLY, MENTOR BECAME AWARE THAT THE LOT NUMBER OF THE SUSPECT MEDICAL DEVICE IS 1003734. ON (B)(6) 2021, MENTOR COMPLETED THE INVESTIGATION ON THE SUSPECT MEDICAL DEVICE. DEVICE EVALUATION SUMMARY: ACCORDING TO THE INFORMATION REPORTED, THE PATIENT DEVELOPED CAPSULAR CONTRACTURE IN THE BREAST. THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE BREAST IMPLANT HAD A CREASE/FOLD ON THE ANTERIOR VIEW AND POSTERIOR VIEW. ADDITIONALLY, TWO TEARS WERE NOTED: TEAR A ON ANTERIOR VIEW WITHIN CREASE AND TEAR B ON POSTERIOR VIEW WITHIN CREASE MEASURING APPROXIMATELY 0.5 CM AND 0.8 CM RESPECTIVELY. MENTOR CONCLUDED THAT THE CAPSULAR CONTRACTURE IN THE PATIENT¿S BREAST WAS THE RESULT OF THE BODY¿S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. IT SHOULD BE NOTED THAT AS PART OF MENTOR¿S QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON 06-MAY-2022, MENTOR BECAME AWARE THAT THE SUSPECT MEDICAL DEVICE WAS MANUFACTURED IN MENTOR¿S MANUFACTURING LOCATED IN LEIDEN, THE NETHERLANDS. IT WAS PREVIOUSLY REPORTED TO THE FDA THAT THE SUSPECT MEDICAL DEVICE WAS MANUFACTURED AT MENTOR¿S TEXAS MANUFACTURING FACILITY. MENTOR MEDICAL SYSTEMS B.V, LOCATED IN LEIDEN, THE NETHERLANDS, IS A FOREIGN MANUFACTURER OF MEDICAL DEVICES THAT ARE NOT CLEARED OR APPROVED FOR SALE IN THE US. IT IS A SEPARATE LEGAL ENTITY FROM MENTOR TEXAS. PER 21 CFR 803.58 AND "MEDICAL DEVICE REPORTING FOR MANUFACTURERS" (FDA GUIDANCE DOCUMENT ISSUED ON NOVEMBER 8, 2016), WE ARE ¿NOT REQUIRED TO SUBMIT MDR REPORTS FOR EVENTS OCCURRING IN OTHER COUNTRIES FOR A DEVICE THAT IS MANUFACTURED IN A FOREIGN COUNTRY AND THAT IS NOT CLEARED OR APPROVED FOR MARKETING IN THE US. HOWEVER, IF MENTOR BECOMES AWARE OF INFORMATION THAT REASONABLY SUGGESTS A DEVICE HAS MALFUNCTIONED AND THAT A SIMILAR DEVICE THAT IS MARKETED IN THE US WOULD BE LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF THE MALFUNCTION WERE TO RECUR, THEN MENTOR SHOULD REPORT THE DEVICE MALFUNCTION THAT OCCURRED OUTSIDE THE US." SINCE MENTOR MEDICAL SYSTEMS B.V. DO NOT MARKET ANY DEVICES IN THE US, MANUFACTURE & MARKET ALL THEIR DEVICES OUTSIDE THE US, AND HAVE NEVER HELD FDA APPROVAL NOR CLEARANCE FOR ANY OF THEIR PRODUCTS, THEIR DEVICES DO NOT MEET THE CRITERIA FOR MDR REPORTABILITY. AS A RESULT, NO FURTHER MDRS WILL BE SUBMITTED FOR THIS EVENT. G1: (B)(4). MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: BILATERAL CAPSULAR CONTRACTURE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) YEAR-OLD FEMALE PATIENT WHO UNDERWENT A BREAST AUGMENTATION PROCEDURE WITH MENTOR MEMORYGEL BREAST IMPLANT 500CC GEL BREAST PROSTHESES DEVELOPED BAKER GRADE III CAPSULAR CONTRACTURE IN BOTH OF HER BREASTS. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION AND REPLACEMENT WITH MENTOR MEMORYGEL BREAST IMPLANT 375CC GEL BREAST PROSTHESES ON (B)(6) 2021. DURING EXPLANTATION SURGERY, IT WAS OBSERVED THAT ONE OF THE REMOVED BREAST PROSTHESES HAD RUPTURED. THIS MEDWATCH REPORT IS FOR THE 2ND OF TWO BREAST PROSTHESES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
743496 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR MEDICAL SYSTEM B.V. 3505001BC 1003734 00081317000341

Patients

Seq Age Sex Outcome Treatment
1 43 YR Female Required Intervention