FDA Adverse Event Malfunction Summary report: N

WM CLARIS CPU

MDR report key: 11850580 · Received May 19, 2021

Report

Report Number
2184149-2021-00169
Event Type
Malfunction
Date Received
May 19, 2021
Date of Event
April 14, 2021
Report Date
November 19, 2021
Manufacturer
ST. JUDE MEDICAL
Product Code
DQK
UDI-DI
05415067009777
PMA / PMN Number
K151911
Removal / Correction Number
2184149-05/05/21-001-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL DATA: G3, H2, H10 CORRECTED DATA: H6

Additional Manufacturer Narrative · 1

THE INVESTIGATION REVEALED THAT AFTER THE WORKMATE CLARIS V.1.2 SOFTWARE UPGRADE, THE WORKMATE CLARIS DWS SCREEN TURNS BLACK/BLANK AND THE USER NEEDS TO REBOOT THE SYSTEM TO REGAIN FUNCTIONALITY.

Description of Event or Problem · 1

DURING THE PROCEDURE, THE DISPLAY SCREENS WENT BLACK. THE DEVICE WAS REBOOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
743688 WM CLARIS CPU COMPUTER, DIAGNOSTIC, PROGRAMMABLE DQK ST. JUDE MEDICAL H700977 7644445 05415067009777

Patients

Seq Age Sex Outcome Treatment
1 Unknown