CASE IV
Report
- Report Number
- 2118418-2008-00001
- Event Type
- Other
- Date Received
- September 29, 2008
- Date of Event
- September 12, 2008
- Report Date
- September 26, 2008
- Manufacturer
- WR MEDICAL ELECTRONICS CO.
- Product Code
- JXE
- PMA / PMN Number
- 910624
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
THERMAL ELECTRIC UNIT, PELTIER (TEU). AN ONSITE VISIT WAS PERFORMED BY PERSONAL TO EVALUATE THE DEVICE AND IT WAS DETERMINED THE THERMAL ELECTRIC UNIT (TEU) WAS DAMAGED CAUSING AN UNEVEN TEMPERATURE GRADIENT ACROSS THE SURFACE. THIS 'HOT-SPOT' MAY HAVE CAUSED THE REPORTED PATIENT BURN AND THIS DEVICE HAS BEEN TAKEN OUT OF SERVICE. THIS WAS CAUSED BY USER DAMAGE TO THE SURFACE OF THE TEU CAUSING A CRACK IN THE CERAMIC SURFACE. USER WAS INSTRUCTED TO USE LIQUID CRYSTAL THERMAL SHEET TO DETECT UNEVEN TEMPERATURE DURING ROUTINE CALIBRATION VERIFICATION OR ANYTIME SURFACE IS SUSPECT. THIS INFORMATION CAN BE FOUND IN THE DEVICE MANUAL.
DURING TESTING AT HOSPITAL, OF A PATIENT THAT IS A FEMALE, WHO DR. REPORTED AS IN GOOD HEALTH AND PART OF AN ONGOING OUTPATIENT TESTING PROGRAM AT THE CLINIC WAS BURNED USING THE CASE IV DEVICE DURING THE HEAT AS PAIN TEST. HE CONFIRMED THAT SHE PARTICIPATED IN CASE IV TESTS IN 2008. AND REPORTED THE BURN ONE WEEK LATER, FROM WHAT DR. SAID, THE BURN ON HER LEFT HAND IS 3/4 CM X 1/2 CM IN A RECTANGULAR SHAPE AND HAS SCABBED OVER. AN ONSITE VISIT WAS PERFORMED BY PERSONAL TO EVALUATE THE DEVICE AND IT WAS DETERMINED THE THERMAL ELECTRIC UNIT (TEU) WAS DAMAGED CAUSING AN UNEVEN TEMPERATURE GRADIENT ACROSS THE SURFACE. THIS 'HOT-SPOT' MAY HAVE CAUSED THE REPORTED PATIENT BURN AND THIS DEVICE HAS BEEN TAKEN OUT OF SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CASE IV | CASE IV | JXE | WR MEDICAL ELECTRONICS CO. | 5000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR | Other |